GAUSSMETERS
SECTION V continued
V
-
B
(
i
)
PROBE ZERO
Remote (Slave) Mode Operation
Command - ZE<channeb<status>
1= Channel 1
2 = Channel 2
3 = Channel 3
must be '0'
(See NOTE)
NOTE: The PROBE ZERO operation will start immediately, so the probe must
be placed in the ZERO GAUSS CHAMBER (or other desired zero field
environment) Before this command is issued. The "statue character must be
'0' in the command string. If the PROBE ZERO operation is successful, the
RETURNED command string (see Section V-D) will also have the "status"
character set to '0.' If an error prevented a successful zeroing of the probe,
"status" will be '1.' See Section IV-I for further information on zeroing errors.
EXAMPLE: "ZE20" zeroes CHANNEL 2's probe.
V-B(j) PROBE RELATIVE
Command - RE<channel><state><select><range><sign><reading>
1 = Channel 1
2 = Channel 2
3 = Channel 3
1
1 = OFF
2 = ON
1 = USE PREVIOUS
2 = GENERATE NEW
Must be
'0000000'
(See NOTE)
NOTE: The PROBE RELATIVE operation will start immediately, so the probe
must be placed in the desired magnetic field environment before this
command is issued. The "range," "sign" and 5-digit "reading characters must
be '0' in the command string. If the "USE PREVIOUS" selection is made, or if
a new relative operation is invoked
and
if the PROBE RELATIVE operation is
successful, the RETURNED command string (see Section V-D) will contain
the following information:
RE<channel><state><select><range>< sign><reading>
I
+
1 = 3 G or 300 µI ( 3 0 mG o r 3 µ T ) "
2 = 30 G or 3 mT ( 3 0 0 mG or 3 0 µT)
3 = 300 G or 30 mT ( 3 G o r 300 µT)
4 = 3 kG or 300 mT ( 3 0 G o r 3 mT)
5 = 30 kG or 3 T ( 3 0 0 G o r 3 0 mT)
6 = 300 kG or 30 T ( 3 kG o r 300 mT)
7 = 3 MG or 300 T
a 5-digit string from
00000 to 29999