![F.W. Bell 9900 Series Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/f-w-bell/9900-series/9900-series_operation-and-maintenance-manual_535142025.webp)
SERIES
9900
DISPLAY FORMAT
CHANNEL-1
BAR
BACK
CHANNEL-2
GRAPH
DIGITS
LIGHT
CHANNEL-3
HELP
ON
ON
ON
MAIN MENU
OFF
OFF
OFF
RUN
9000
SERIES
GAUSSMETERS
SECTION I V continued
Menu Operation/Gaussmeter
Programming
Either the digital reading or the bargraph, or both, can be turned off to suit the
user's requirements. By turning
both
off, the user eliminates the channel
from the screen and the channel and its probe is placed in a standby position.
If only the digital reading is turned off, the reading's polarity ('-, or + or —) will
remain on the screen as an indication of the type of field (ac or dc) being
measured. This also disables the ability to gather field readings in the
"SLAVE" mode of remote operation, discussed in Section V.
If ambient lighting is inadequate to clearly see the display during the
measurement process, the backlight can be turned on for better visibility.
This is done by choosing any of the channel selections.
Figure IV-F
DISPLAY FORMAT Menu
IV-H SETUP S AV E -LOAD
A "SETUP" is the condition of the gaussmeter (range settings, display format,
classifiers, etc.) while the instrument is operating in the MEASURE mode.
The gaussmeter can permanently store up to six separate machine set-ups.
The purpose of SETUP SAVE-LOAD is to allow the user to program a setup
for a particular application, for instance, measuring a batch of permanent
magnets. This setup can be saved and later retrieved, instantly programming
the gaussmeter without the need to re-enter all of the parameters via the
MENU mode each time the gaussmeter is used. It should be remembered
that the present machine SETUP is saved when the instrument is powered
down and will be reinitialized to that SETUP when powered up again. See
Section III-I for further information.
The six SETUPs are titled "A" thru "F." When REVIEW, SAVE or LOAD is
chosen, the cursor will advance to the title column. At this point the up (11 or
down (I) pushbuttons are used to select the desired SETUP title. At the
same time that title's SETUP information block will appear on the right-hand
side of the screen, allowing the user to view the programming information.
Similar to the three-channel MEASURE mode display screen, the top
information block is for CHANNEL 1, the middle CHANNEL 2 and the lower
CHANNEL 3.
The setup is then validated by the ENTER pushbutton.
In the SAVE mode
the PRESENT machine setup will replace the setup that appears on the
screen.
Care should be used in this mode to avoid unintentionally overwriting
a setup. In the LOAD mode, the setup for the selected title
will replace the
present machine setup.
REVIEW simply allows the user to view the
information without affecting the setup.
20