SERIES
9900
9000
SERIES
GAUSSMETERS
SECTION I
I
-A INTRODUCTION
GENERAL
DESCRIPTION
The SERIES-9900 gaussmeter represents the latest developments in the
science of measuring magnetic flux density using the Hall effect. The unit is
available in three configurations capable of processing a single channel
(Model 9901), two channels (Model 9902) or three channels (Model 9903) of
magnetic field information. Either steady-state (dc) or alternating (ac) fields
can be measured. When matched with the appropriate F.W. Bell fourth-
generation Hall generator probe, fields as low as 10 pG (0.0011.1T) or as high
as 2.9999 MG (299.99 T), at frequencies up to 50 kHz, can be measured with
extreme accuracy and 4-3/4 digit resolution.
The SERIES-9900 features PEAK HOLD, AUTORANGING, CLASSIFIER and
RELATIVE operation, auto ZEROing, GAUSS or TESLA readout, digital and
bargraph representation, diagnostics and remote operation with an IEEE-488
(GPIB) instrumentation bus and an RS-232 communications port. All
information is displayed on an illuminated graphics liquid crystal display (LCD).
The gaussmeter employs a menu-driven format to allow the user to program
all aspects of gaussmeter operation with ease and speed. Each gaussmeter
channel is completely and independently programmable.
I-B FUNCTIONAL DESCRIPTION
Figures I-A and I-B depict the SERIES-9900 front and rear panels,
respectively. The gaussmeter is housed in a standard 6.87" high x 17.67"
wide x 16.18" deep cabinet featuring pop-up feet for tabletop use and
(optional) brackets for rack mounting.
FRONT PANEL
(1) DISPLAY
(2) POWER SWITCH
(3) ZERO GAUSS CHAMBER
(4) "T" /CHANNEL 1 "RESET"
PUSHBUTTON
5
1920-pixel (240 x 64) dot matrix (graphics)
LCD with backlight. Contrast is preset at the
factory.
Rocker type power switch with international
legends ("0"=OFF, ''1"=ON).
Recessed double-wall Mu-metal chamber
used to shield the probe from stray magnetic
fields during the PROBE ZERO operation.
Attenuation is 80 dB in a 300 G field.
Dual-function momentary-contact
pushbutton switch used to reset PEAK
readings for CHANNEL 1 in the MEASURE
mode or advance the cursor up in the MENU
mode.