METALJET D2+
INSTALLATION
Revision 19, November 2016
Page 4-6
The electron gun/x-ray head and the pump box must be properly fastened in order to avoid dangerous
situations where personnel may get injured or equipment may get damaged.
The electron gun/x-ray head should be properly fastened using the six M6-threaded holes underneath
the base plate, see Figure 4-5. Mount the base plate onto a support with dimensions smaller than the
base plate to avoid that the side covers take up the weight of the x-ray head. The center-of-gravity for
the x-ray head is indicated in Figure 4-4.
The pump box can either be rack mounted or positioned directly on the floor. Use the side brackets on
the front of the pum
p box to attach the pump box to a 19” rack, see
Figure 4-6, or to fasten it relative
to the floor
. Note that the 19” rack must in turn also be stable in its
relative position to the electron
gun/x-ray head.
Figure 4-4.
Location of center-of-gravity for the x-ray head.