20
WT.050.580.001.DE.IM.0714
Installation
Sensor measuring modules for MFC
4.
4.5
Inserting and connecting the sensors
Please note
Before installing the sensors, please note Chapter 4.2
Measurement Inputs in the Instruction Manual „MFC”.
Observe the max. back pressure (pressurized version). Please
consult the membrane sensor data sheet for this figure. You can
request this figure from Evoqua Water Technologies GmbH or
locate it in the Instruction Manual „MFC”.
Arrangement of
the sensors
A Membrane sensor: FC1, CD7, OZ7, TC1, TC1-S
B Redox
C Fluoride or conductivity
D pH
1
Remove the protection caps from the sensors.
2
Install sensors (see figure above) in the cell body cover.
The sensors are marked as follows:
Membrane sensor for free chlorine, chlorine dioxide, ozone and
total chlorine (A)
mV:Sensor for Redox, marked "mV" (B)
pH:Sensor for pH value, marked "pH" (C)
µS:Sensor for conductivity, marked "LF325" (D)
A
D C B D C B
DEPOLOX
®
5 DEPOLOX
®
5
non-pressurised version pressurised version
C B D D B A
VariaSens non-pressurised VariaSens pressurised
version
version
A
C
Содержание Wallace & Tiernan W3T158761
Страница 1: ...WALLACE TIERNAN SENSOR MEASURING MODULES AND RETROFIT KITS FOR MFC INSTALLATION MANUAL...
Страница 2: ...2 WT 050 580 001 DE IM 0714 Sensor measuring modules for MFC...
Страница 4: ...4 WT 050 580 001 DE IM 0714 Contents Sensor measuring modules for MFC...
Страница 26: ...26 WT 050 580 001 DE IM 0714 Installation Sensor measuring modules for MFC 4...
Страница 36: ...36 WT 050 580 001 DE IM 0714 Index Sensor measuring modules for MFC 6...
Страница 37: ...Sensor measuring modules for MFC 37...