Sensor measuring modules for MFC
Installation
4.
15
4.2
Installing and wiring plug-in card 1 to 7
1
Remove the slot's dummy cover.
Save for later use!
2
Insert and screw in the plug-in card.
3
If necessary, wire the sensor cable terminal strip and insert
into the plug-in card.
4
Attach the module identificaiton (I) on the cable according to
the module slot (mod 1 = 1, mod 2 = 2, mod 3 = 3 and
mod 4 = 4).
The respective wiring diagrams are listed in Chapter 5. “Wiring
Diagram" on page 27 or in the Instruction Manual „MFC” "Wiring
Diagrams".
"DES" module for
DEPOLOX
®
5
and/or
DEPOLOX
®
4
and
PT100
Terminal assignment:
1: Red (WRK)
4: White (REF)
5: Blue (CNT)
Shield: Black
Depending on the module slot, the included plug must be used for
connection on the left or right side.
Terminal assignments for plug-in card with PT100:
"pH" module
Please note
The sensor cable is already wired with pH plug-in card!
"mV" module
Please note
The sensor cable is already wired with mV plug-in card!
"F
-
" module
Please note
The sensor cable is already wired with F
-
plug-in card!
Содержание Wallace & Tiernan W3T158761
Страница 1: ...WALLACE TIERNAN SENSOR MEASURING MODULES AND RETROFIT KITS FOR MFC INSTALLATION MANUAL...
Страница 2: ...2 WT 050 580 001 DE IM 0714 Sensor measuring modules for MFC...
Страница 4: ...4 WT 050 580 001 DE IM 0714 Contents Sensor measuring modules for MFC...
Страница 26: ...26 WT 050 580 001 DE IM 0714 Installation Sensor measuring modules for MFC 4...
Страница 36: ...36 WT 050 580 001 DE IM 0714 Index Sensor measuring modules for MFC 6...
Страница 37: ...Sensor measuring modules for MFC 37...