6. Operation
DEPOLOX® 400 M
62
EN
6.3.7 ”Information” menu
Various system information, e.g. the installed
software version, is described in this menu.
6.3.8 ”User administration“ menu
The electronics module offers the option of
using up to three user levels. Different rights
are assigned to each user level. The colored
user symbol in the Home screen shows on
what level the user is logged in.
Fig. 2
Home menu
A Actual user adminstration level
No symbol
•
User logged out.
•
Read rights for all settings and change of
operation mode.
User symbol white = Level 1
•
User logged in on Level 1.
•
Read rights for all settings, sensor cali-
bration and change of user.
User symbol yellow = Level 2
•
User logged in on Level 2.
•
Read rights for all settings, sensor cali-
bration, change of operation mode,
change limit values and setpoints and
change date and time.
Connection - RS485 interface
Function:
Off or RS485 WT
Selection of the RS485 Interface func-
tion.
The RS485 Interface supports bus com-
munication with the Process Monitoring
System or other higher-level systems
that support the RS485-WT protocol.
RS485-WT protocol (new address refe-
rence list)
Busadress:
Bus address setting at the RS485 Inter-
face (RS485 WT) 00 to31
Status:
RxD, TxD
Aktive error
Backup and Reset
Reset to factory setting:
Under Factory setting, the device can be
reset to the factory settings. When the
factory settings have been restored, the
reset settings have to be entered again.
You can select which settings are to be
reset.
Reset measurement & control parameters:
Reset system settings:
Reset user administration:
Reset dosing average:
NOTICE
To reset the selected setting to the factory
setting, press the “Execute” button.
NOTICE
The user administration of the web visuali-
zation via Internet browser differs from the
user administration on the display of the
electronics module (chapter 6.3.9).
NOTICE
No symbols are displayed also with disabled
access control; in this case, write access to
all settings is possible.