6. Operation
DEPOLOX® 400 M
60
EN
6.3.6 ”Settings” menu
The device settings not relating to measured
values are configured in the Settings menu.
1 Call up the HOME menu.
2 Tap the symbol.
3 Tap
the
Settings Menu.
4 Enter desired settings.
NOTICE
The control function for the DEPOLOX®
400 M is only to be used for constant flow
applications and is only an on/off type of
controller. Under no circumstances should
this control function be used in applications
where the process water flow changes for
any reason.
System
Setup
Setup - Controller
Disinfection controller:
On/Off
Enables or disables the disinfection con-
troller (only to be used for constant flow
application)
Setup - Measurement
Desinfection:
On/Off
Enables or disables pH measurement
and controller
pH/OPR/Fluoride:
On/Off
Enables or disables pH, ORP (Redox) or
fluoride measurement
Gerneral information
System name:
Freely definable
Language:
German/English/French/Italian/
Croatian/Dutch/Japanese
Hold Function:
On/Off
The hold function is used to either buf-
fer all measured values or keep them
constant during calibration. This pre-
vents the output of invalid control sig-
nals by the sensor during the calibration
process and also the output of erratic
values from measurements via mA-sig-
nal and communication interfaces. The
function is enabled when the calibration
menu is opened and disabled when the
menu is closed.
Display
Brightness:
0 to 100 %
Screensaver:
Off, 30 s, 1/5/15/30 min, 1 h
Color scheme:
Design 1 to 5
Calibrate LED:
This setting can be used for white
balance of the LED glow stick color if
color deviations occur.
Red, yellow, blue: 50 to 100 %
Time/Date - Date
Time:
00:00 to 24:00 / 00:00 to 12:00
24 h Format:
On/Off
Safety
Sample water delay time:
00:00 to 10:00 min
The sample water delay time determi-
nes the time after which dosing is deac-
tivated, e.g. in the event of sample
water Stop. While the delay time is run-
ning, “Const.” is displayed in the Auto-
matic symbol.
Feed delay time:
00:00 to 120:00 min
The feed delay time delays the start of
dosing when the device is switched on,
after switch-on, when the operating
mode has been changed, after Control-
ler STOP. The rundown of the selected
time can be interrupted by selecting the
“Start now!” button.