![Evoqua DEPOLOX Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/evoqua/depolox/depolox_instruction-manual_2447564027.webp)
DEPOLOX® 400 M
3. Description
27
EN
Membrane sensor
CD7 (chlorine dioxide)
Membrane sensor
OZ7 (ozone)
Measured value
Chlorine dioxide, selective toward
Cl
2
, Br
2
, H
2
O
2
, cross-sensitivity to
O
3
, peracetic acid
Ozone, selective toward Cl
2
, Br
2
,
H
2
O
2
, cross-sensitivity to ClO
2
,
peracetic acid
Power supply
un12 – 15 VDC, 11 mA
Typical output signal
approx. 10 µA pre 1 mg/l (ppm)
ClO
2
approx. 10 µA pre 1 mg/l (ppm)
O
3
Measuring system
Membrane-covered 2-electrode system
Reference electrode
Silver / silver halide / potassium halide solution
Working electrode
Gold
Further materials
PVC, silicone rubber, stainless steel
External dimensions
Ø 25 mm (1"), length 175 mm (6.9")
Connector-cable
combination
Cable length 1.2 m
Electrolyte
Diluted potassium halide solution,
100 ml bottle, store in a dark
place at 59 to 77°F (15°C to 25°C),
can be used for a max. of 2 years if
still evenly clear
Diluted potassium halide solution,
100 ml bottle, store in a dark
place at 59 to 77°F (15°C to 25°C),
can be used for a max. of 1 year if
still evenly clear
Range
0,05 to 20 mg/l (ppm) ClO
2
0,02 to 10 mg/l (ppm) O
3
Analysis
DPD 1 + DPD 3
Zero point calibration
Not necessary (zero point signal at
0 mg/l ClO
2
= 0 µA)
Not necessary (zero point signal at
0 mg/l O
3
= 0 µA)
Response time
T
90
< 20 sec.
T
90
< 50 sec.
Influence of the pH
value
No signal influence up to stability
limit Cl0
2
No signal influence up to stability
limit Ozon
Temperature com-
pensation
Internal temperature compensation 41 to 113°F (5 – 45°C)
Storage temperature
14 to 113°F (-10 – +45 °C) (without electrolyte)
Max. pressure
1.5 bar (21.76 PSI) with suitable flow cell, no pressure surges and/or
vibrations. Observe the back pressure of the flow cell!
Water quality
All types of water, swimming pool,
drinking, service and process
water (limescale deposits can
block the membrane).
All types of water, swimming pool,
drinking, service and process
water (also containing surfac-
tants), (limescale deposits can
block the membrane).
Conductivity
> 1 µS/cm to max. 40 mS/cm
Flow
1.6 to 9.2 gal/hr (6 – 35 l/h), as constant as possible
Service life
Service life of the electrolyte in operation approx. 6 months, service
life of the membrane cap typically 1 year (less with poor water qua-
lity, e.g. sand, dirt)
Cross-sensitivity
Ozone, peracetic acid
Chlorine dioxide, peracetic acid
Selectivity
Toward chlorine, bromine, hydrogen peroxide