Eversense CGM User Guide
Calibration Stability Agreement
The table below compares the percentage of sensor glucose values to the YSI reference at various time points
after a calibration entry. As an example, in the table below, 88% of the Eversense values were within 15 mg/dL
(for reference readings of 80 mg/dL or less), and within 15% (for reference readings over 80 mg/dL) of the
reference value 8 to 10 hours after a calibration entry.
Table 16 – Eversense Calibration Stability Agreement in the PRECISION Study
Percent of CGM System Readings Within
Time
from
Calibration
Number of
Paired
CGM-YSI
Percent
15/15%
of
Reference
Percent
20/20%
of
Reference
Percent
30/30%
of
Reference
Percent
40/40%
of
Reference
> 40/40%
of
Reference
0 - 2 Hours
4,034
86.0%
93.6%
98.0%
99.3%
0.7%
2 - 4 Hours
3,979
85.6%
92.8%
98.4%
99.5%
0.5%
4 - 6 Hours
2,308
84.3%
92.2%
97.7%
99.0%
1.0%
6 - 8 Hours
1,614
84.3%
92.7%
97.8%
99.4%
0.6%
8 - 10 Hours
1,372
88.0%
94.4%
98.5%
99.6%
0.4%
10 - 12 Hours
1,295
86.1%
92.9%
98.1%
99.2%
0.8%