19
181
Eversense CGM User Guide
Alert Setting
(mg/dL)
Confirmed
Event Detection
Rate
Missed
Detection
Rate
True Alert
Rate
False Alert
Rate
Low
Alert
60
76%
24%
81%
19%
70
90%
10%
95%
5%
80
94%
6%
96%
4%
90
96%
4%
95%
5%
High
Alert
120
99%
1%
96%
4%
140
99%
1%
96%
4%
180
98%
2%
95%
5%
200
98%
2%
95%
5%
220
98%
2%
95%
5%
240
98%
2%
95%
5%
300
93%
7%
94%
6%
Table 11 – Eversense High and Low Glucose Alert Performance (Threshold Only) in the PRECISION Study
The table below shows an assessment of the ability of the Eversense CGM System to detect high and low glucose
levels, and assess true positive alerts vs false positive alerts. These are shown as a percentage of alerts provided
compared to the YSI reference values at various thresholds.
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 181
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 181
2/26/20 12:59 PM
2/26/20 12:59 PM