Eversense Alert Performance
The tables in this section show an alert performance assessment. The Confirmed Event Detection Rate shows the
percentage of time the Eversense CGM System confirmed the reference value by presenting an alert within a 15 minute
window of a reference value beyond the alert setting threshold. The Missed Detection Rate shows the percentage of
time the Eversense CGM System did not present an alert within a 15 minute window of a reference value beyond the
alert setting threshold. The True Alert Rate shows the percentage of time the alert from the CGM system was confirmed
by a reference value within a 15 minute window of the alert being presented. The False Alert Rate shows the percentage
of time the alert from the CGM system was not confirmed by a reference value within a 15 minute window of the alert
being presented.