Eversense CGM User Guide
Alerts and Notifications Displayed on the Apple Watch
(continued)
Apple W
at
ch
Displa
y
Description
Calibrate Now
Appears to alert you that your calibration is due.
If you do not calibrate within 4 hours, glucose
values will no longer be displayed.
Sensor Check
Appears once when the system’s internal checks
detect instability with the sensor which requires a
return to calibration Initialization Phase.
A
ctions
Confirm the alert on your Eversense CGM
Mobile App.
Tap
Calibrate
on your mobile
device to enter a calibration value.
Confirm the alert on your Eversense CGM
Mobile App.
In the Initialization Phase, you
must perform 4 fingerstick calibrations spaced
2 - 12 hours apart. Display of glucose readings
will resume after the 2nd successful fingerstick
calibration.