Error Messages 

Instant S

The test strip may be damaged, not 
properly inserted, or was previously used. 
Remove and reinsert the test strip, or 
replace it if damaged or previously used.

A meter or test strip error has occurred.
Repeat the blood glucose test. If a 
second E-3 code appears, perform a 
control test with the control solution 
and a new test strip.

• If the control result is within the acceptable range, 
review the proper testing procedure and repeat the 
blood glucose test with a new test strip.
• If the control result is not within the acceptable range, 
see the Understanding Out-of-Range Control Results 
section in the chapter Control Tests in the User’s 
Manual.
This error could appear if the cap on the test strip 
container was not closed tightly. The test strips may have 
been damaged due to improper storage or handling.
In rare cases, an E-3 error code may indicate that your 
blood glucose is extremely high and above the system’s 
measuring range. If you do not feel like your blood 
glucose is extremely high, repeat the blood glucose test. 
See the Unusual Blood Glucose Results section in the 
chapter Blood Glucose Tests in the User’s Manual.
If the E-3 code still appears for your blood glucose test, 

contact your healthcare professional immediately.

Not enough blood or control solution was 
drawn into the test strip for measurement 
or was applied after the test had started.
Discard the test strip and repeat the blood 
glucose or control test.

Blood or control solution was applied to the 
test strip before the flashing drop symbol 
appeared on the display.
Discard the test strip and repeat the blood 
glucose or control test.

An electronic error occurred.
Remove the battery, press and hold the 
meter button for at least 2 seconds, and 
reinsert the battery. Perform a blood glucose 
or control test.

The temperature is above or below the 
proper range for the system.
Refer to the test strip package insert for 
system operating conditions. Move to an 
area with the appropriate conditions and 
repeat the blood glucose or control test.  
Do not artificially heat or cool the meter.

The battery is almost out of power.
Change the battery now. If the message 
reappears after the battery has been 
replaced, remove the battery again, press 
and hold the meter button for at least  
2 seconds, then reinsert the battery.

The meter’s internal clock may need 
to be reset.
Press the meter button. If the time and 
date were set, you will need to reset 
them. See the Time and Date section in 
the chapter Your New System.

The test strip may be damaged.
Repeat the blood glucose or control 
test with a new test strip.

Your blood sample may contain a high 
level of ascorbate.
Contact your healthcare professional.

Fluid or foreign material may be present 
in the test strip slot.
Remove and reinsert the test strip or 
repeat the blood glucose or control 
test with a new test strip. If the error 
persists, contact Roche.

An electronic error has occurred.
Contact Roche.