background image

5

cartrIdge InStallatIon and actIVatIon

Remove the outer shrink-wrap and protective cap from the 

1. 

2SR-BW & MR-600 RO cartridges. Position a bucket under the 

2SR -BW cartridge head. Slowly open the Inlet shut-off valve. 

Apply power to the system by plugging the power cord into the 

receptacle. Within moments, the display water should illuminate 

and the system will go into PRESSURIZE mode. Once a solid flow 

of water has been established, flush 1 gallon. Unplug power cord. 

Empty bucket.
Install 2SR-BW cartridge. Align the lugs on the new cartridge 

2. 

with the slots in the head. Holding the head and cartridge firmly, 

press the cartridge upward into the head fully, and then turn the 

cartridge to the right until it stops.
Disconnect 3/8" OD line from 2SR-BW to Booster Pump inlet at 

3. 

Booster Pump. Route line into bucket and secure. Plug-in power 

cord. Once a solid flow of water has been established, flush 1 

gallon. Unplug power cord. Re-connect 3/8" OD line to Booster 

Pump inlet. Empty bucket.
Route a blue line from the 3/8" storage tank drain valve to drain 

4. 

temporarily. Open storage tank drain valve.
Install MR-600 cartridge. Align the lugs on the new cartridge 

5. 

with the slots in the head. Holding the head and cartridge firmly, 

press the cartridge upward into the head fully, and then turn the 

cartridge to the right until it stops.
Plug-in power cord. Once a solid flow of water has been 

6. 

established, flush 5 gallons from storage tank drain (about 10 

minutes). Unplug power cord. 

WARNING

  do not uSe tHIS Product Water! It may contain 

a preservative solution from the ro cartridge and 

should not be consumed!

recoVery adjuStment

Measure the raw water TDS. Refer to Table 1 on page 6. Find 

1. 

the TDS range in the column labeled “Raw Water TDS” that 

corresponds to your TDS measurement. Read across that row 

to determine maximum recovery, permeate to reject ratio and 

permeate/reject volumes.  
Using 2 - 32 oz. or 1,000 ml graduated containers, apply power 

2. 

to the system and collect the permeate flow (from storage tank 

drain valve) in one container and the reject (WASTE) flow in the 

second container. Compare the volumes collected to the data 

shown in the table. Adjust the reject needle valve to increase or 

decrease the flow. Empty the containers and collect additional 

samples. 
Once the ratio of Permeate-to-Reject closely matches the data 

3. 

in Table 1, the Reject Only volume requires measurement. Using 

only 1 graduated cylinder, collect the reject flow for 1 minute.

note:

  Reject volume in 1 minute may exceed 32 oz./1,000 ml. If 

so, use a larger container, or collect samples in multiple 

containers and add all collected volumes together. 

Unplug power cord. Close storage tank drain valve.

4. 

Return to page 6, Table 2. Find the volume range in the column 

5. 

labeled “Reject Volume per Minute” that corresponds to the 

volume collected in Step 4. Read across that row to identify the 

correct capillary tube combination. 
Locate the correct capillary tubes from the parts kit. Remove the 

6. 

2 - 1/4"OD tubes from the upper Y-connector. Insert the capillary 

tubes into each line and re-assemble to the Y-connector. 

note:  

In some cases, the capillary tube requires trimming. Use a 

SHARP utility knife to cut the tube cleanly without tearing or 

crushing the tube.

Remove Reject Needle valve assembly and reconnect 3/8" tube. 

7. 

Plug-in power cord. Allow system to run and fill tank. This will 

take 5-10 minutes.
Refer to Table 2 again. Under the column labeled “Flush Time in 

8. 

Seconds”, find the value that corresponds to the Raw Water TDS. 

Enter this value in the controller under “FLUSH TIME X1”. Adjust 

“FLUSH INT X2” to 1.5 Hours.

figure 5.

 RO Reject Flow Control Assembly

TUBE

FLOW

CONTROL

Capillary 

Not Trimmed

TUBE

FLOW

CONTROL

Capillary

Trimmed

Содержание ENVI-RO

Страница 1: ...increases the inlet pressure to maintain consistent permeate production plus eliminates back pressure on the permeate output This innovative approach to both boosting and eliminating back pressure al...

Страница 2: ...ler Power Supply Storage Tank Wall Mount Bracket 2 MR 600 RO cartridge 3 2SR BW Cartridge 4 Parts Kit 5 Inspect the carton for damage Report any damage to freight carrier immediately and retain all pa...

Страница 3: ...ectionable taste Plastic pipe or reinforced tubing are generally very good choices for RO water distribution materials The size of the product water tubing piping should be 3 8 ID minimum Distances of...

Страница 4: ...alue in seconds is entered for this setting Range is 0 995 seconds in 5 second increments Use the keys to adjust this value Press Enter to save value and go to next screen Flush Int X2 This parameter...

Страница 5: ...o your TDS measurement Read across that row to determine maximum recovery permeate to reject ratio and permeate reject volumes Using 2 32 oz or 1 000 ml graduated containers apply power 2 to the syste...

Страница 6: ...1 321 67 9 679 751 1000 30 0 1 to 2 33 30 0 300 70 0 700 Table 2 Capillary Tube and Flush Time Selector Reject Volume per Minute Capillary Tube Flush Time in Seconds Ounces Milliliters 1 2 0 0 6 1 0 1...

Страница 7: ...odor is detected from 8 the water exiting the storage tank or connected equipment Placing ENVI RO Into Service Once the cartridges have been activated the storage tank and distribution piping sanitiz...

Страница 8: ...pm Dependant on Pressure volume in storage line sizes Volume Gallons Dependant on Pressure volume in storage Pressure typical 60 100 psi 4 1 6 9 bar Salt Rejection 90 minimum Recovery Varies based on...

Страница 9: ...9 maintenance section Performance Log Test Date Outlet Flow ml min Drain Flow ml min Inlet Pressure Temp psi F Inlet TDS ppm STEAMER Outlet TDS ppm COFFEE Outlet TDS ppm Reduction NOTES...

Страница 10: ...capacity if applicable c Low quality RO water 3 Refer to problem 1 a Refer to problem 1 a RO cartridge failure b Replace RO cartridge b Low water pressure at water 4 using equipment Tubing piping run...

Страница 11: ...13 EV312827 RESTRICTOR BLUE 735 ml min 13 EV312828 RESTRICTOR WHITE 1052 ml min 13 EV312829 RESTRICTOR GREY 1577 ml min 13 EV312830 RESTRICTOR RED 125 ml min 13 EV312831 RESTRICTOR BROWN 189 ml min 1...

Страница 12: ...instructions on the MR cartridge label Place a bucket under the cartridge to catch any excess water that may spill from the head Remove the old cartridge by turning the cartridge to the left until it...

Страница 13: ...13 13 6 346 23 583 406 16 32 Side View Front View Top View 813 Figure 7 Dimensional Drawing...

Страница 14: ...14 Figure 8 General Installation Arrangement...

Страница 15: ...15 NOTES...

Страница 16: ...00 942 1153 www everpure com Printed in U S A EV3128 25 Rev A SE08 For sales replacement components and service contact your Everpure dealer or Everpure LLC 1040 Muirfield Drive Hanover Park Illinois...

Отзывы: