![Eurotron IRtec Rayomatic 20 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/eurotron/irtec-rayomatic-20/irtec-rayomatic-20_instruction-manual_2440980008.webp)
________________________________________________
_______________________________________________
Instruction Manual MM850341 ed.01a
___________________________________________8____________________________________________
1.1.2
IRtec Rayomatic 40
•
Measuring ranges:
see par 1.2.2 Table A
•
Accuracy:
@ 23°C ±5°C and
ε
=1.0
IRtec Rayomatic 40-100:
±0.5% of rdg.
IRtec Rayomatic 40-160:
±0.5% of rdg.
IRtec Rayomatic 40-814:
±1% of rdg. or ±1°C whichever is greater
IRtec Rayomatic 40-390:
±1% of rdg.
IRtec Rayomatic 40-460:
±1% of rdg.
IRtec Rayomatic 40-510:
±1% of rdg.
IRtec Rayomatic 40-343:
±1% of rdg. or ±3°C whichever is greater
IRtec Rayomatic 40-790:
±1% of rdg. or ±2°C whichever is greater
•
Repeatability:
IRtec Rayomatic 40-100:
±0.25% of rdg.
IRtec Rayomatic 40-160:
±0.25% of rdg.
IRtec Rayomatic 40-814:
±0.5% of rdg. or ±0.5°C whichever is greater
IRtec Rayomatic 40-390:
±0.5% of rdg.
IRtec Rayomatic 40-460:
±0.5% of rdg.
IRtec Rayomatic 40-510:
±0.5% of rdg.
IRtec Rayomatic 40-343:
±0.5% of rdg. or ±1.5°C whichever is greater
IRtec Rayomatic 40-790:
±0.5% of rdg. or ±1.5°C whichever is greater
•
Response time:
IRtec Rayomatic 40-100: 28ms (t95)
IRtec Rayomatic 40-160: 28ms (t95)
IRtec Rayomatic 40-814: 100ms (t95)
IRtec Rayomatic 40-390: 100ms (t95)
IRtec Rayomatic 40-460: 100ms (t95)
IRtec Rayomatic 40-510: 100ms (t95)
IRtec Rayomatic 40-343: 1000ms (t95)
IRtec Rayomatic 40-790: 1000ms (t95)
IRtec Rayomatic 40-ratio: : 38ms (t95)
•
Type of detector:
IRtec Rayomatic 40-100:
Silicium
IRtec Rayomatic 40-160:
InGaAs
IRtec Rayomatic 40-814 :
Thermopile
•
Spectral band:
see par 1.2.2 Table A
•
Emissivity:
adjustable by PC from 0.30 to 1.00
•
Temperature drift:
±0.1 °C/°C for the band exc18 to +28°C
•
Working temperature:
from -20 to +60 °C (without cooling)
from 0 to +50 °C (for laser operation)
•
Storage temperature:
from -30 to +70 °C Silicium and InGaAs detectors
from +10 to +60 °C Thermopile
•
Digital communication:
Bell 202 superimposed on 2-wire current loop
RS232 with optional adapter
•
Output signal:
4/20 mA 2 wire current loop – max load 700
Ω
•
Target pinpointing:
optional on-board laser aiming system with remote command
•
Environmental rating:
IP65 (NEMA-4)
•
Power supply:
Loop: from 12 to 32 Vdc
Laser pinpointing: 12 to 32 Vdc
•
Dimension / Weight:
ø 45 mm x 200 mm - 0.5 Kg