Instruction Manual MM850341 ed.01a
___________________________________________22____________________________________________
6
INSTALLATION
IRtec Rayomatic
installing is very easy. The cylindrical aluminium chassis is externally threaded (M44x1.5) and
allows quick connecting with all available accessories.
The following procedure could help you for installing your
IRtec Rayomatic
thermometer.
•
Configure the thermometer. Set the emissivity on the
IRtec Rayomatic 20
moving the 2 rotative switches.
Connect the
IRtec Rayomatic 40
to the PC and set the mathematical function, the emissivity and the
temperature range by software.
•
Connect, if necessary, all accessories (air purge, support, sighting tube, etc.). Fix the IRtec Rayomatic to the
process.
•
Connect the signal cable from your data acquisition system to the instrument terminals. Pay attention to the
polarity.
•
Aim accurately the thermometer to your target. Refers to Target vs. distance table on section 1. Be sure that
your target is bigger than the measuring area. The optional laser pinpointing can help you for best
installation.
IMPORTANT
DO NOT
PULL OUT THE ELECTRONIC BOARD FROM THE CYLINDER
. I
T COULD BE DIFFICULT TO REASSEMBLE THE
THERMOMETER
.
WARNING
DO NOT
INSTALLING THE THERMOMETER IN ROOM WITH TEMPERATURE HIGHER THAN
+60°C
WITHOUT AN ADEQUATE
COOLING SYSTEM
.
6.1
Electrical connection
The temperature signal is converted, amplified and linearized by the internal electronics that makes available a
4/20 mA linear output.
Eurotron
supply a connector for 4/20 mA active current loop connection.
The unit is supplied complete with a shielded multi-wires electrical cable.
The following procedure could help you for electrical cable substitution on your IRtec Rayomatic thermometer:
1. Remove the 3 screw on the flange on the rear of the instrument.
2. Open the cable gland and insert the shielded signal cable (
φ
3.9 – 6.5 mm) through the flange.
3. Unsheathe the cable and prepare the wire for connection.
4. Unscrew partially the cable clamp. Insert correctly the wires. Pay attention to the polarity. Screw the cable
clamp.
5. Put near the flange to the cylinder and screw firmly.
6. Close the cable gland.
7. Insert the 3 screw to the flange holes and accompain them manually. Use an adequate screw driver. Look up
(1 or 2 rotation) alternatively the 3 screw.
IRtec Rayomatic 20 connections:
Laser Power
Supply
Output
4/20mA
100.0
-
+
IN 4/20mA
Indicator with 24Vdc internal power supply