MULTIPOINT CIRCULAR-CHART RECORDER INSTALLATION AND OPERATION MANUAL
Page 4 - 23
HA250385
Issue 4 Feb 99
4.9.1 EMBEDDED SEQUENCES (Cont.)
EXAMPLES
An alarm going active on channel 3 has 'Print Message 1 on going active' as one of its jobs.
If Message 1 were set up to be:
<TIME><.TA><-PV>
then the current time and the tag and process value of chan-
nel 3 would be printed on the chart.
If, instead, Message 1 were
<TIME><6.TA><6-PV>
then the current time and the tag and process value of chan-
nel 6 would be printed on the chart when the channel 3 alarm went active.
4.10 INTERNAL EVENTS
As standard, there are six internal events, which can be triggered by one or more sources, and which can generate up
to two jobs each when active. Input sources can be ANDed or ORed, so multiple logical inputs can be used.
For example, to send a log to the chart when any alarm on channel 1 and channel 2 and channel 4 are active, we
could set up events 1 and 2 as follows:
Event 1:
Event 2:
Job 1:
Enabled
Enabled
Log 1 to chart
Source 1(S1) AND Source 2 (S2)
S1 AND S2
On going active.
S1:Alm on channel 1.
S1:Alm on channel 4
S2:Alm on channel 2.
S2:Event 1
4.10.1 Event sources
Event sources are:
Alarm on Ch N (any Alarm on specified channel)
Chart is online (See section 4.4.3 - Stop after One Revolution).
Clock failure
Event N (Another specified event - see figure above)
Glb Channel alarm (Alarm on any channel)
Glb UnAck Ch Alm (Unacknowledged alarm on any channel)
Operator Key (See sections 3.6 and 4.7 - Operator action)
Power up
System error (Section 3.8)
Ch1
S1
S2
Event 1
S1
S2
Event 2
Ch 2
Ch 4
Event Job 1
Send log 1
to chart
S1.S2
S1.S2