SanaBike 500 easy
Article-Nr.: 2.520132c
Revision: 10/2018
Page 15 of 21
Operation modes
Remote operation
When the load is automatically controlled by an exercise/training program, this is done by the ergometry
control program of an external master device such as PC, ECG unit or similar, depending on the application.
7.1.1
Requirements for remote operation
For running an ergometry program which is controlled by an external training program, the following applies
•
the ergometer needs to be individually adjusted to the patient, as described in chapter
Fehler! Ver-
weisquelle konnte nicht gefunden werden.
.
•
the ergometer needs to be connected to the master device via an interface (RS232, WLAN, Bluetooth).
(Note: For patient safety, the RS-232 interface is galvanically isolated.)
•
the transmission rate (Baud rate) needs to be set correctly.
•
the data protocol type P10 needs to be selected (ergometer and master device).
Manual operation
In addition to the standard remote operation, the load can also be defined manually by the user in steps of
5 watt on the ergometer, which is described in this chapter.
7.2.1
Requirements for manual operation
For performing an ergometry manual, the ergometer should be justificated individually to the patient acc.
section
Fehler! Verweisquelle konnte nicht gefunden werden.
.
7.2.2
Change into manual operation mode
Switching to manual operation mode and back to remote operation is done via the standard display
„
ergometry screen
“
. By pressing the display in the range of the current load value, the
„Manu
al Operation
“
screen will be displayed to start and manually adjust the load.
Icon for switching
the operation modes
Standard display
„
ergometry screen
–
remote operation
”