SanaBike 500 easy 

 

 

 

Article-Nr.: 2.520132b 

Revision: 10/2018 

Page 12 of 21 

 

 

Adjusting the saddle height and handle bars 

The  saddle  can  be  adjusted  upwards  or  downwards  for  optimal  exercise  position.  It  can  be  adjusted  for 
heights between 120 and 210 cm. 

T-handles that protrude from under both the handlebar and saddle pipe are used for adjustment (for manual 
saddle  height  adjustment).  After  loosening  the  respective  T-handle,  the  saddle  rod  can  be  moved  up  or 
down to the desired height. When the adjustment is finished, the clamp needs to be tightened again. 

The  saddle  clamp  is  designed  for  patients  weighing  up  to  200  kg.  To  attain  optimal  clamping  action,  only 
moderate strength is needed to tighten the clamps. It is recommended to adjust the T-handles in clamped 
position,  with  the  handle  positioned  vertically  downward.  The  clamp  can  be  unlocked  by  pulling  back  the 
handle  from  its  lock.  Once  the  clamp  is  unlocked,  it  can  be  turned  to  the  desired  position  and  simply  re-
leased, after which it will lock itself again automatically. Use this position of the clamp for further reference 
to ensure secure clamping. 

In the same way, use the clamp to optimally position the ergonomic handle bar. 

 

 

 

Содержание SanaBike 500 easy

Страница 1: ...Article Nr 2 520132b Revision 10 2018 Page 1 of 21 Bike ergometer SanaBike 500 easy User guide...

Страница 2: ...elevant requirements of the EC medical devices directive 93 42 EEC Annex I This declaration loses its validity if the above devices are modified without ergosana s consent ergosana GmbH s quality mana...

Страница 3: ...ssembling 10 3 2 2 Connecting 10 UNIT COMPONENTS 11 Displayscreens 11 Setting the language 11 Adjusting the saddle height and handle bars 12 SAFETY NOTES 13 Precautions during operation use 13 Safety...

Страница 4: ...18 Messtechnische Kontrollen MTK und Sicherheitstechnische Kontrollen STK 18 Checking and setting the supply voltage 18 Changing a mains fuse 18 Eliminating electromagnetic interferences 18 Disposal 1...

Страница 5: ...optionally electrical adjustment Standardised saddle pipe change of saddle possible at any time Impact and scratch resistant casing easy to clean High performance control electronics Touch display sho...

Страница 6: ...the case of arrhythmogenic right ventricular disease coronary artery disease Exercise trials for patients with frequency adaptive pacemaker systems to define the ideal intervention frequency Proof of...

Страница 7: ...ure 10 mmHg with signs of myocardial ischaemia angina pectoris ST elevation 0 1 mV horizontal ST depression 0 2 mV or lacking systolic increase in blood pressure respectively blood pressure 240 mmHg s...

Страница 8: ...s 7 Running gear 8 Clamp for handle bar adjustment 9 180 rotatable console with touchscreen display and rpm display Accessories Every device comes with Mains cable User manual Inspection report Potent...

Страница 9: ...gometer in steps of 5 watts Display Touch display 57x43 mm Power supply 230 VAC with 50 Hz or 115 VAC with 60 Hz The unit is suitable for use in electric networks according to CISPR 11 group 1 class B...

Страница 10: ...the clamp lever downwards after fixing the saddle in place To do so pull the clamp outwards position and then release With the help of the base adjustors on the lower rear side of the ergometer adjus...

Страница 11: ...creens A foil covered touchscreen 57 x 43 mm is located at the front of the console Depending of the range of functions of the ergometer the display shows on to three different screens which shows all...

Страница 12: ...adjustment is finished the clamp needs to be tightened again The saddle clamp is designed for patients weighing up to 200 kg To attain optimal clamping action only moderate strength is needed to tight...

Страница 13: ...n the pedal straps The holding straps on the pedals must fit perfectly across the upper side of the shoe and be fastened with a Velcro strap For patients who weigh more than 140 kg or are taller than...

Страница 14: ...an be adjusted see chapter Fehler Verweisquelle konnte nicht gefunden werden the following needs to be observed Adjusting the saddle position The saddle needs to be in a horizontal position To set the...

Страница 15: ...is galvanically isolated the transmission rate Baud rate needs to be set correctly the data protocol type P10 needs to be selected ergometer and master device Manual operation In addition to the stan...

Страница 16: ...transmitted from the ergometer to the master device but commands of the master device will be not accepted 7 2 3 Starting adjusting and terminating manual operation mode The manual load definition wil...

Страница 17: ...be treated with the disinfectant Incidin Plus fluid as needed 8 2 1 Disinfection interval It is recommended to perform a disinfection in the following situations Daily commissioning After use with tra...

Страница 18: ...ans After the device is disconnected from the power supply open the power supply cover on the bottom of the unit The voltage can then be changed on the power supply board with a special voltage select...

Страница 19: ...ecommunication devices senders and the ergometer SanaBike 500 depending on the output performance of the communication device as indicated below Recommended minimal distance between the telecommunicat...

Страница 20: ...aBike 500 easy Article Nr 2 520132b Revision 10 2018 Page 20 of 21 Characteristics for the braking moment control Family of characteristics for the area of the braking moment control 999 Speed Load Wa...

Страница 21: ...ial ist dealers if your device requires servicing If this is not possible please contact the company s central service department Ergosana service department ergosana GmbH Truchtelfinger Str 17 D 7247...