background image

 

SanaBike 500 easy 

 

 

Article-Nr.: 2.520132c 

Revision: 10/2018 

 

Page 11 of 21 

 

 

Unit components 

The console mounted on the handle bar is rotatable by 180°. Depending on the application, it can be facing 
either the examiner or the user. 

At the front, a high-resolution touchscreen provides all important information at a glance as well as all ele-
ments to control and program the ergometer. 

 At the top of the console, a 7-segment display shows the current revolutions per minute of the pedals. 

 

 

Displayscreens 

A foil-covered touchscreen (57 x 43 mm) is located at the front of the console. Depending of the range of 
functions of the ergometer, the display shows on to three different screens, which shows all important er-
gometry data and allows to the control the functions of the ergometer. Moreover, when performance is 30 
watt or higher, a red arrow indicates if the correct range of revolutions is exceeded. 

 

Setting the language 

The  device  is  delivered  with  the  selected  language;  however,  the  language  can  be  changed  by  a  service 
technician at any time. The following languages are available. 

 

German 

 

English 

 

French 

 

Italian 

 

Spanish 

 

 

Ergometry screen 

Indication to correct  

revolutions per minute

 

Содержание SanaBike 500 easy

Страница 1: ...Article Nr 2 520132b Revision 10 2018 Page 1 of 21 Bike ergometer SanaBike 500 easy User guide...

Страница 2: ...elevant requirements of the EC medical devices directive 93 42 EEC Annex I This declaration loses its validity if the above devices are modified without ergosana s consent ergosana GmbH s quality mana...

Страница 3: ...ssembling 10 3 2 2 Connecting 10 UNIT COMPONENTS 11 Displayscreens 11 Setting the language 11 Adjusting the saddle height and handle bars 12 SAFETY NOTES 13 Precautions during operation use 13 Safety...

Страница 4: ...18 Messtechnische Kontrollen MTK und Sicherheitstechnische Kontrollen STK 18 Checking and setting the supply voltage 18 Changing a mains fuse 18 Eliminating electromagnetic interferences 18 Disposal 1...

Страница 5: ...optionally electrical adjustment Standardised saddle pipe change of saddle possible at any time Impact and scratch resistant casing easy to clean High performance control electronics Touch display sho...

Страница 6: ...the case of arrhythmogenic right ventricular disease coronary artery disease Exercise trials for patients with frequency adaptive pacemaker systems to define the ideal intervention frequency Proof of...

Страница 7: ...ure 10 mmHg with signs of myocardial ischaemia angina pectoris ST elevation 0 1 mV horizontal ST depression 0 2 mV or lacking systolic increase in blood pressure respectively blood pressure 240 mmHg s...

Страница 8: ...s 7 Running gear 8 Clamp for handle bar adjustment 9 180 rotatable console with touchscreen display and rpm display Accessories Every device comes with Mains cable User manual Inspection report Potent...

Страница 9: ...gometer in steps of 5 watts Display Touch display 57x43 mm Power supply 230 VAC with 50 Hz or 115 VAC with 60 Hz The unit is suitable for use in electric networks according to CISPR 11 group 1 class B...

Страница 10: ...the clamp lever downwards after fixing the saddle in place To do so pull the clamp outwards position and then release With the help of the base adjustors on the lower rear side of the ergometer adjus...

Страница 11: ...creens A foil covered touchscreen 57 x 43 mm is located at the front of the console Depending of the range of functions of the ergometer the display shows on to three different screens which shows all...

Страница 12: ...adjustment is finished the clamp needs to be tightened again The saddle clamp is designed for patients weighing up to 200 kg To attain optimal clamping action only moderate strength is needed to tight...

Страница 13: ...n the pedal straps The holding straps on the pedals must fit perfectly across the upper side of the shoe and be fastened with a Velcro strap For patients who weigh more than 140 kg or are taller than...

Страница 14: ...an be adjusted see chapter Fehler Verweisquelle konnte nicht gefunden werden the following needs to be observed Adjusting the saddle position The saddle needs to be in a horizontal position To set the...

Страница 15: ...is galvanically isolated the transmission rate Baud rate needs to be set correctly the data protocol type P10 needs to be selected ergometer and master device Manual operation In addition to the stan...

Страница 16: ...transmitted from the ergometer to the master device but commands of the master device will be not accepted 7 2 3 Starting adjusting and terminating manual operation mode The manual load definition wil...

Страница 17: ...be treated with the disinfectant Incidin Plus fluid as needed 8 2 1 Disinfection interval It is recommended to perform a disinfection in the following situations Daily commissioning After use with tra...

Страница 18: ...ans After the device is disconnected from the power supply open the power supply cover on the bottom of the unit The voltage can then be changed on the power supply board with a special voltage select...

Страница 19: ...ecommunication devices senders and the ergometer SanaBike 500 depending on the output performance of the communication device as indicated below Recommended minimal distance between the telecommunicat...

Страница 20: ...aBike 500 easy Article Nr 2 520132b Revision 10 2018 Page 20 of 21 Characteristics for the braking moment control Family of characteristics for the area of the braking moment control 999 Speed Load Wa...

Страница 21: ...ial ist dealers if your device requires servicing If this is not possible please contact the company s central service department Ergosana service department ergosana GmbH Truchtelfinger Str 17 D 7247...

Отзывы: