ELite 5, User Manual
Page 106 of 159
13.3 External Devices
The External Devices panel allows entry of external LIS settings. The ‘ELite 5’ analyzer
support external LIS communications over a serial communications link using the Erba
Lachema version 3.1 protocol. It also supports an Ethernet communications link over the HL7
version 2.5 protocol.
The External Devices panel lets you change the following settings:
•
Sending port baud rate – select 9600 or 115200 baud for serial LIS connection.
•
Automatic LIS – The ‘ELite 5’ will transmit every measurement result as soon as it is
available if this check box is checked.
•
LIS – the ‘ELite 5’ analyzer will use an Ethernet connection rather than a serial
connection of this check box is checked.
•
IP – the IP address of the external LIS computer should be entered here if an
Ethernet LIS connection is used. IP addresses consist of four numbers from 0 to 255
separated by periods. See your IT administrator for more information about the IP
address of your host LIS computer.
•
Port – the port number of the external LIS computer should be entered here with nor
commas if an Ethernet LIS connection is used. Ports are a single number between 0
and 65535. See your IT administrator for more information about the port number of
your host LIS computer.
•
Bidirectional LIS – check this check box if the LIS computer will download lists for one
of the list modes for automatic sample processing.
Figure 78. External Devices Settings Panel
Содержание ELite 5
Страница 1: ...Erba Lachema s r o Kar sek 2219 1d 621 00 Brno CZ ELite 5 Hematology Analyzer USER MANUAL INS00068...
Страница 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Страница 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Страница 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Страница 143: ...ELite 5 User Manual Page 143 of 159...
Страница 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...