4.2.6 Left Side Assembly
To access the left or right side assemblies, the side panels must be removed. Please refer to
section 15.1.2 for information about removing the side panels.
Pressure Sensors
Small Measure Buffer
Drain Buffer,
Mix, RBC & WBC
Chambers
Valves 1-22
Valves 23-44
Diluent Buffers
Diff-5p, Lyse-5p Buffers
Sheath Buffer
Figure 13. Left Side Assembly
•
Pressure sensors measure the differential pressures of vacuum buffer for
impedance base and optical measurement and empty chamber process.
•
Small Measure buffer is a vacuum buffer used for impedance measurement.
•
Sheath buffer is a reservoir for Erba•Dil-5P diluent for optical measurements.
•
Diff-5P buffer is a reservoir for Erba•Diff-5P reagent.
•
Lyse-5P puffer is a reservoir for Erba•Lyse-5P reagent.
•
Diluent buffers are reservoirs for Erba•Dil-5P diluent.
•
Valves 1-44 connect pneumatic components to carry out dilution, reaction,
measurement, and waste disposal processes.
•
Drain buffer controls chamber draining procedures.
•
Mix chamber performs the first step of RBC dilution.
•
RBC chamber performs the second step of RBC dilution and contains the RBC
aperture.
•
WBC chamber performs the WBC dilution and holds the WBC aperture and HGB
photometric measurement system.
Содержание ELite 5
Страница 1: ...Erba Lachema s r o Kar sek 2219 1d 621 00 Brno CZ ELite 5 Hematology Analyzer USER MANUAL INS00068...
Страница 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Страница 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Страница 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Страница 143: ...ELite 5 User Manual Page 143 of 159...
Страница 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...