20.1 Reagent Consumption
Function
Reagent consumption / function (ml)
Erba Lyse-5P
Erba Diff-5P
Erba Dil-Diff
Start Up
13.6
1.7
128
Measure Blank
5.7
0.7
52
Measure 5 part
5.7
0.7
52
Measure Calibration
5.7
0.7
52
Measure QC
5.7
0.7
52
Measure SSM blank
5.7
0.7
60
Measure SSM 5-part
5.7
0.7
60
Measure SSM Calibration
5.7
0.7
60
Measure SSM QC
5.7
0.7
60
Standby
0
0
11.2
Wakeup
0
0
10
Standby in SSM
0
0
11.2
Wakeup in SSM
0
0
11
Cleaning
6.7
0.7
133
Hard cleaning
6.7
0.7
144.2
Clean in SSM
6.7
0.7
138
Hard clean in SSM
6.7
0.7
154
Flow cell cleaning
0
0
63
Shear Valve cleaning
9
1.2
100.2
Offline (overnight) rinsing
6.7
0.7
72.2
**Prime Diluent (full)
0
0
103
**Prime Lyse
7.5
0
2.5
**Prime Diff5P
0
4
0
**Prime all
22
11.5
103
*DrainDilu
0
0
120
*DrainLyse
60
0
0
*DrainDiff5P
0
60
0
*DrainAll
60
60
180
FillUp
22
12
105
Shutdown
7.7
0.7
153.2
*
If internal chambers are filled (volume of chambers: 50 ml each, diluent 2x50 + 50 sheath)
**
Prime functions fill internal chambers. The following pneumatic function uses reagents from the inside puffer while the
inside puffer is continuously refilled to a steady level.
Содержание ELite 5
Страница 1: ...Erba Lachema s r o Kar sek 2219 1d 621 00 Brno CZ ELite 5 Hematology Analyzer USER MANUAL INS00068...
Страница 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Страница 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Страница 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Страница 143: ...ELite 5 User Manual Page 143 of 159...
Страница 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...