ELite 5, User Manual
Page 124 of 159
The shear-valve is in contact with the sample blood. Wear gloves while
cleaning the shear-valve. Handle the materials use to clean the shear-valve
as potentially infectious material.
The parts which should be disassembled / re-assembled are fixed with so called thumb-
screws. No screwdriver or other similar tools are required to tighten these screws.
Go to the Maintenance menu, and select Shear Valve Clean function to initiate the
procedure.
Figure 94. Maintenance menu-shear valve
The ‘ELite 5’ asks for confirmation to start the procedure. After clicking or tapping the ’OK’
button the ‘ELite 5’ empties the shear-valve and the connecting tubes. A small amount of
liquid may remain inside the tubing and inside the shear-valve.
When the procedure completes, the ‘ELite 5’ prompts the user to proceed the shear valve
cleaning with the following message:
Figure 95. Shear valve cleaning info
Don’t press the ‘OK’ until the shear valve has been cleaned and re-assembled again!
Perform the shear valve cleaning according to the instructions below.