enware AQUABLEND ESQX ATAB C1L3 Скачать руководство пользователя страница 18

18 

Call 1300 369 273 

www.enware.com.au

IMAGE 55

IMAGE 56

IMAGE 57

IMAGE 58

ºC (-)

 

CLOSE

ºC (+)

 

OPEN

THERMAL DISINFECTION PROCEDURE

STEP 6

To prepare the valve for hot water flush, it is necessary to make 

the solenoid stay in the open position. To do this first activate 

 

the sensor to open solenoid. As soon as a click from solenoid  

is heard and solenoid opens, disconnect the solenoid cable  

from sensor at the connector. The solenoid will stay in the open  

position until later when it is reconnected to the sensors.  

SEE IMAGES 55 & 56

STEP 7

With the front plate off, temporarily install the spout back on so 

the water can flow into the basin. SEE IMAGE 57

STEP 8

Open the integral isolation valve for the hot water supply, while 

keeping cold side closed. (

NOTE

: hot water will not flow yet, 

due to cartridge position being in OFF position.) 

STEP 9

Prepare for hot water to flow out of outlet, taking precautions 

 

to address the risk of scalding from the hot water flowing out 

 

of the outlet.

STEP 10

Pasteurisation or heat decontamination procedure can now 

be carried out according to the methods stated in the relevant 

standards and regulations.

To start flow of hot water, slowly turn the thermostatic mixer 

cartridge to OPEN position using 3/8” Allen key. SEE IMAGE 58

WARNING

: Full temperature hot water will flow out of outlet.

Maximum hot water temperature allowed for the valve is 70°C 

 

for hot water flush, due to limitation of the solenoid valve and 

spout aerator.

STEP 11

Once decontamination procedure has completed, turn the 

thermostatic mixer cartridge to OFF position, and turn off hot 

water supply by closing the isolation valve.

STEP 12

Unscrew flexible hose and bypass adaptors. Re-fit Strainer / 

Check Valve Assemblies to ports

STEP 13

Connect sensor to solenoid.

STEP 14

Restore hot and cold water supplies by opening the integral 

isolation valves.

STEP 15

Take spout off. Then fit front plate and install spout back on.

STEP 16

Commission the valve and set the valve to required temperature, 

according to the Commissioning Procedure on page 19.

Содержание AQUABLEND ESQX ATAB C1L3

Страница 1: ...are Australia Pty Limited 9 Endeavour Rd Caringbah NSW 2229 Australia Ph 61 2 8556 4000 info enware com au I00057_May 19 Installation and User Guide AQUABLENDTM ESQXTM TOUCH FREE THERMOSTATIC MIXERS A...

Страница 2: ...PAGE 8 INSTALLATION PROCEDURE FRONT OF WALL PAGE 10 USER GUIDE PAGE 15 THERMAL DISINFECTION PROCEDURE PAGE 17 COMMISSIONING PAGE 19 SERVICE AND MAINTENANCE PAGE 21 CHANGING THE SENSOR PROGRAM PAGE 28...

Страница 3: ...G1 2 BSP WARM OUTLET DENOTES 310 X 190 FACE PLATE POSITION 166 BLOCKOUT 98 C C 180 222 7 7 95 127 5 6 THRU TYP 73 3 106 73 3 106 105 8 138 5 180 8 213 5 190 FLANGE DIMENSION 310 FLANGE DIMENSION G1 2...

Страница 4: ...above outlet On Demand Sensor Front of Wall Component Only 230mm Fixed Spout with Laminar Flow Mains Powered ATWM C2L3 8Lpm ATWM C2L5 6Lpm ATWM C2L6 4 5Lpm MAINS AUTO SENSE Sensor below outlet Auto Se...

Страница 5: ...T MAINS NOTE WHEN PLACING AN ORDER IN WALL back of wall component is ordered separately to FRONT OF WALL component Battery operated IN WALL back of wall component MUST be paired with battery operated...

Страница 6: ...ommended that the hot and cold water supply pressures be balanced within 10 Min 20kPa Max 500kPa Static Inlet Pressures For testing purposes system commissioning Max 1000kPa Hot Temperature Supply Ran...

Страница 7: ...ure that the system operating conditions fall within the recommended operating range specified in Technical Information Installation Conditions To ensure that the thermostatic mixer operates correctly...

Страница 8: ...he desired location for box support Enware recommend 13mm ply wood fixed between two vertical in wall studs NOTE plywood needs to cover the entire back section of the box so the valve body can be fixe...

Страница 9: ...into a power outlet and place the end connector of the 12V transformer in the box via the cable access hole SEE IMAGE 6 Ensure at least 50mm of cable is available in the box to connect to the sensor...

Страница 10: ...s to be trimmed so it finishes flush with finished wall face Discard cover SEE IMAGE 8 Check that no part of the box protrudes past the finished wall and deburr trimmed edges STEP 2 Fit the chrome bac...

Страница 11: ...STEP 4 Cut back the 1 2 outlet thread so you are left with 15mm protruding past the support bracket SEE IMAGE 12 STEP 5 Apply thread sealant or tape to outlet thread and screw on the spout retainer u...

Страница 12: ...sing the grub screw SEE IMAGES 14 15 STEP 7 Commission the valve as stepped out in commissioning procedure on page 19 NOTE Do not high pressure test the supply system as this may damage the thermostat...

Страница 13: ...g your hand in front of sensor about 10cm away for 1 second for On Demand Sensor models or by holding your hand in front of sensor for Auto Sense models Listen for the solenoid to click open Water is...

Страница 14: ...Allen key to secure it in place SEE IMAGE 23 24 Ensure the grub screw fits into the groove of the spout retainer and is fitted all the way into the spout STEP 16 Take a 3 8 Allen key insert into the t...

Страница 15: ...ON the tap runs for at least 2 seconds Intelligent Afterflow Function During this time the tap cannot be turned off Maximum continuous flow period is 2 minutes OPTIONAL SENSOR FUNCTIONS Afterflow can...

Страница 16: ...elligent Afterflow Function During this time the tap cannot be turned off If tap is not turned OFF it will automatically turn OFF after a set period Factory default setting is 15 sec range 8 sec 180 s...

Страница 17: ...egral isolation valves located within each inlet connector SEE IMAGE 50 STEP 3 Using a 3 8 Allen key unscrew the strainer check valve assembly from both sides and keep at hand SEE IMAGES 51 52 STEP 4...

Страница 18: ...scalding from the hot water flowing out of the outlet STEP 10 Pasteurisation or heat decontamination procedure can now be carried out according to the methods stated in the relevant standards and reg...

Страница 19: ...led servicing In all cases the following must be checked to ensure correct operational performance of the SQX valve The intended installation matches the performance brief of the SQX point of use ther...

Страница 20: ...e by closing the integral isolation valve on the cold inlet side The outlet flow should quickly cease flowing As a rule of thumb the flow should be less than 0 1Lpm following the isolation Restore the...

Страница 21: ...or in place STEP D Remove the faceplate carefully over the spout connector Annual servicing can be done with faceplate bracket in place STEP 2 Isolate the hot and cold supplies to the Aquablend eSQX T...

Страница 22: ...e mixer SEE IMAGES 28 29 STEP 2 The check valve strainer assembly should be cleaned with a dilute water solution of suitable descaling solvent such as CLR checked for physical damage and then thorough...

Страница 23: ...MAGES 33 34 STEP 3 Use an appropriate size spanner or wrench unwind the thermostatic cartridge out of the mixer body Lower the cartridge down into the base of the box until it is free of the mixer bod...

Страница 24: ...noid membrane Enware part ENMS212 The membrane and lower cap should be cleaned with a diluted water solution of suitable descaling solvent such as CLR checked for physical damage and then thoroughly r...

Страница 25: ...in the sensor lens Activate the sensor and check that the solenoid clicks when the sensor is activated STEP 6 Re install the battery into the box Re assemble the faceplate dress flange and spout STEP...

Страница 26: ...E 46 STEP 4 Insert measurement probe into Pete s Plug and open isolation valves Check for leaks SEE IMAGE 47 STEP 5 Operate the mixer and conduct in service tests as required ACCESS TO INSPECTION PORT...

Страница 27: ...cing upright SEE IMAGE 49 SERVICING THE SENSOR Enware Product Code ENMS236 Auto Sense 6V Battery Op Sensor ENMS237 On Demand Sensor 6V Battery Op Sensor ENMS238 Auto Sense 12V Mains Power Sensor ENMS2...

Страница 28: ...nctions to suit different hand wash applications and procedures STEP 1 Disconnect power to sensor STEP 2 Reconnect power and wait 5 seconds The red LED light in the sensor lens will turn on for 3 seco...

Страница 29: ...Automatic on sensor range max 170mm 4s 1 second afterflow Sensitivity level 2 20 170mm 4S 1s 2 2 mins 4 Automatic on sensor range max 250mm 4s 1 second afterflow Sensitivity level 2 20 250mm 4S 1s 2 2...

Страница 30: ...7 Automatic off On Demand Sensor active with 180 second runtime Sensor range 20 100mm 20 100mm 4s 1s 1 180 sec 0 sec 8 Automatic off On Demand Sensor active with 15 second runtime Sensor range 20 100m...

Страница 31: ...authority 2 The valve manufacturer supplier requirements 3 The Australian Standards for Plumbing and Drainage YES NO YES NO YES NO If NO give details and action taken PRINT ALL DETAILS or MARK WITH A...

Страница 32: ...Date of next service due Previous service carried out by Date of previous service Valve installed by Date of installation Temperature range of warm water at outlet Neonatal and children 38 40 C Adult...

Страница 33: ...contains debris or is damaged Clean the cartridge ensuring that all debris is removed and components are not damaged Replace if necessary Strainers contain debris Clean strainers ensuring debris is r...

Страница 34: ...tain debris Clean strainers ensuring debris is removed Either full hot or cold flowing from outlet fixture The temperature cam is incorrectly set Re set temperature Hot cold water has migrated to othe...

Страница 35: ...incorrectly connected Replace sensor and battery A new battery typically lasts between 3 to 5 years depending on frequency of use Sensor red light constantly blinks Low voltage Replace battery Battery...

Страница 36: ...n of that system must be in accordance with its manufacturer s recommendations the Code and all relevant statutory and local State or Territory requirements 4 The valve must be returned to ENWARE with...

Отзывы: