The EA800 base unit is normally locked to prevent unauthorized use. The currently active function of the
F1
soft
key (
UNLOCK
or
LOCK
) is displayed above the key. The locked and unlocked states are described below.
Note:
The base unit locks automatically after 30 minutes of inactivity if the user does not press the
LOCK
soft
key (
F1
).
■
Locked: This is the default state and limits access to the EA800 to viewing only.
UNLOCK
is displayed above the
F1
soft key indicates that the base unit is
currently locked. Pressing the
F1
soft key prompts the user to enter a password
to unlock the base unit, permitting full access to all screens and functions. The
HOME
key,
MENU
key, and the
MAIN MENU
options shown on the screen at right
are available to the user when the base unit is locked.
■
Unlocked: When the base unit is unlocked,
LOCK
displayed above the
F1
soft key, and all information,
programming, and maintenance screens may be viewed and programming changes may be made. Pressing
the
F1
soft key locks the EA800 base unit and protects it from unauthorized or unintended programming
changes, log downloads, and firmware uploads.
The base unit allows the use of two passwords for unlocking:
■
The factory default password is
0800
. This password cannot be changed or deleted.
■
A second, optional password may be set by the user. To create a second user password, see “Changing the
Password” on page 63.
J6
INPUT 1
Wired input for Relay 1 external temperature, water, 4-20mA, contact closure,
or humidity sensor.
INPUT 2
Wired input for Relay 2 external temperature, water, 4-20mA, contact closure,
or humidity sensor.
INPUT 3
Wired input for Relay 3 external temperature, water, 4-20mA, contact closure,
or humidity sensor.
INPUT 4
Wired input for Relay 4 external temperature, water, 4-20mA, contact closure,
or humidity sensor.
J8
OUTPUT 1
Form C relay alarm output for Relay 1.
OUTPUT 2
Form C relay alarm output for Relay 2.
OUTPUT 3
Form C relay alarm output for Relay 3.
J9
OUTPUT 4
Form C relay alarm output for Relay 4.
OUTPUT 5
Form C relay alarm output for Relay 5.
OUTPUT 6
Form C relay alarm output for Relay 6.
J10
OUTPUT 7
Form C relay alarm output for Relay 7.
OUTPUT 8
Form C relay alarm output for Relay 8.
AUX OUT
Form C relay output that activates upon an alarm from any of the relays
(Relay 1 through Relay 8).
J13
Antenna
RF receive and transmit
J14
Antenna
RF receive and transmit
Table 3
EA800 Base Unit Connector Functions — continued
Connector
Designation
Function
Содержание Alert EA800
Страница 1: ...D 011 0152 Electronic Multi Zone Environmental Alarm System Installation Owner s Manual ...
Страница 20: ...16 D 011 0152 Relay Operation ...
Страница 74: ...70 D 011 0152 Exporting the Stored Logs ...
Страница 81: ...Appendix A Screen Maps D 011 0152 77 Figure 19 Sensors Screen Map See Figure 20 See Figure 21 ...
Страница 84: ...80 D 011 0152 Figure 23 Data Log Screen Map ...
Страница 85: ...Appendix A Screen Maps D 011 0152 81 Figure 24 System Screen Map Dependent on format See Figure 25 ...
Страница 86: ...82 D 011 0152 Figure 25 Configuration Screen Map ...
Страница 88: ...84 D 011 0152 ...
Страница 90: ...86 D 011 0152 ...
Страница 96: ...92 D 011 0152 ...