Combination Blanket/Fluid Warmer
Blanket Chamber
Fluid Chamber
Front of
warmer
Zone 1
(fluid, left)
Zone 2
(fluid, right)
Zone 2
(blanket shelf,
top right)
Zone 1
(blanket,
top left)
Zone 4
(blanket,
bottom right)
Zone 3
(blanket,
bottom left)
Service
Part numbers and drawings are subject to change without notice.
How to Replace the Fuse
WARNING
For protection against fire and electrical shock
use only UL listed 20A, 250V fast acting fuses,
6.3mm x 32mm (F1, F2). Access should be made
by qualified service technicians only.
A hospital grade cord must be used. Refer to the
owner’s manual or contact the manufacturer
for acceptable cords. The warmer must be
connected to an equivalent receptacle marked
“hospital grade”.
1 . Unplug the power cord
1
from the wall outlet and
the power inlet
2
on the warmer .
2 . Unscrew the cap
3
on the fuse holder .
3 . Pull the fuses
4
out of the fuser holder .
4 . Replace fuse
4
with a new fuse .
5 . Screw the cap
3
back onto the fuse holder .
1
4
3
2
Zone Heating Element Locations