11
MN-39362 •
Rev 0 • 04/18
• Titan Series
EC1850BL
Combination Blanket/Fluid Warmer
Using the Blanket Controller
Temperature Range:
32°C-71°C (90°F-160°F)
1
5
6
7
2
3 4
How to Set the Temperature Scale
1 . While the blanket controller is in the off mode,
press and hold the temperature recall button
1
for
four (4) seconds .
2 . Press the up arrow button
3
or the down arrow
button
5
to change to Fahrenheit (°F) or Celsius (°C) .
How to Start the Warmer
1 . Connect the warmer to an appropriate hospital
grade receptacle — specifi ed on the electrical
information page .
2 . Locate the circuit breaker on the back of the warmer .
3 . Push the circuit breaker to the ON (I) position .
4 . Press the ON/OFF button
4
. The previous temperature
set-point is displayed
2
.
How to Set the Operational Sound Volume
1 . While the blanket controller is on, press the
temperature recall button
1
and the down arrow
button
5
simultaneously . The current sound volume
setting is displayed
2
.
2 . Press the up arrow button
3
or the down arrow button
5
to change the volume . Volume settings range from
0 (mute) to 12 (loud) .
NOTE:
The alarm volume is set at the maximum (12) and
can not be disabled.
How to Display the Compartment
Temperature
1 . Press the temperature recall button
1
.
The temperature at the compartment sensor is
displayed
2
for fi ve (5) seconds . Then the blanket
chamber temperature set-point is displayed .
How to Lock the Controller
The controller can be locked to prevent changes being
made to the temperature set-point or the mode selection .
1 . Press the ON/OFF button
4
and the up arrow button
3
simultaneously . The lock icon
7
illuminates .
How to Unlock the Controller
1 . Press the ON/OFF button
4
and the down arrow
button
5
simultaneously . The controller unlocks and
the lock icon
7
goes off .
How to Mute the Alarm
When the controller senses an alarm condition (see
troubleshooting guide), the controller sounds an alarm
and the alarm icon
6
fl ashes .
1 . Press the ON/OFF button
4
to acknowledge and mute
the alarm .
NOTE:
If the alarm continues or reoccurs, the warmer needs
to be serviced.
Power Fail Detection
The controller stores all settings and resumes operation
using these settings in the event of a power failure . When
the power is restored, the controller sounds an alarm
once . The LED screen
2
fl ashes .
1 . Press the ON/OFF button
4
.
The LED screen
2
stops fl ashing .
NOTE:
The controller displays the length of time of the
power outage in hours and minutes (hh:mm) for
five (5) seconds. Then the controller displays the set
operating temperature.
2 . Make sure the product inside the warmer is at the
correct temperature before using .
Operation Instructions