14
MN-39362 •
Rev 0 • 04/18
• Titan Series
EC1850BL
Combination Blanket/Fluid Warmer
Operation Instructions
Using the Fluid Controller
1
7
8
9
10
11
2
4
3
5 6
How to Lock the Controller
The controller can be locked to prevent changes being
made to the temperature set-point or the mode selection .
1 . Press the ON/OFF button
6
and the up arrow button
5
simultaneously . The lock icon
11
illuminates .
How to Unlock the Controller
1 . Press the ON/OFF button
6
and the down arrow
button
7
simultaneously . The controller unlocks and
the lock icon
11
goes off .
How to Mute the Alarm
When the controller senses an alarm condition (see
troubleshooting guide), the controller sounds an alarm
and the alarm icon
10
fl ashes .
1 . Press the ON/OFF button
6
to acknowledge and mute
the alarm .
NOTE:
If the alarm continues or reoccurs, the warmer needs
to be serviced.
Power Fail Detection
The controller stores all settings and resumes operation
using these settings in the event of a power failure . When
the power is restored, the controller sounds an alarm
once . The LED screen
2
fl ashes .
1 . Press the ON/OFF button
6
.
The LED screen
2
stops fl ashing .
NOTE:
The controller displays the length of time of the
power outage in hours and minutes (hh:mm) for
five (5) seconds. Then the controller displays the set
operating temperature.
2 . Make sure the product inside the warmer is at the
correct temperature before using .
Temperature Range:
• Irrigation (IRR) temperature range:
32°C to 66°C (90°F to 150°F)
• Injection (INJ) temperature range:
32°C to 40°C (90°F to 104°F)