7
MN-39362 •
Rev 0 • 04/18
• Titan Series
EC1850BL
Combination Blanket/Fluid Warmer
Electrical Information
Locate the Rating Tag
Verify the power requirements for the appliance . The power specifi cation is located on the
appliance identifi cation rating tag . This tag is permanently attached to the appliance .
Hazardous
Voltage Present
CAUTION
Power source must match voltage identified
on appliance rating tag. The rating tag
provides essential technical information
required for any appliance installation,
maintenance or repairs. Do not remove,
damage or modify the rating tag.
DANGER
To prevent serious personal injury, death,
or property damage:
Do not use this warming appliance in
the presence of flammable anesthetic
mixtures (with air or with oxygen or
nitrous oxide).
Not category AP or APG equipment
Safety Class I
Equipment
Protective Earth
Ground Symbol
Medical Equipment classifi ed by Underwriters
Laboratories with Respect to Electrical Shock,
Fire and Mechanical Hazards only, in Accordance
with UL 61010–1 and CAN/CSA C22 .2 No . 61010–1 .
E471516
Grounding reliability can only be achieved when appliance is
connected to an equivalent receptacle marked “Hospital Grade .”
Wire diagram is located under top lid of appliance.
Power Requirements
Dimensions (HxWxD)
With casters:
74 .8" x 30 .0" x 28 .0" (1900mm x 762mm x 712mm)
Capacity
Fluid compartment:
30 1-liter bottles or
24 1-liter bags
Blanket compartment: 30-40 blankets (16 .4 ft
3
)
Weight**
Net: 457 lbs (207 kg)
Ship: 568 lbs (258 kg)
Shipping dimensions: 38" L x 38" W x 90" H
Electrical
120 VAC — 50/60 Hz, 1 ph
1 .3 kW, 11 .0 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-20P
20A - 125V Plug
Hospital Grade
IP-X0
E471516
220 VAC — 50/60Hz, 1 ph
1 .2 kW, 5 .5 Amps
Type B Equipment
CEE 7/7*
220-230V Plug
IP-X0
230 VAC — 50/60 Hz, 1 ph
1 .3 kW, 5 .6 Amps
Type B Equipment
BS 1363 Plug*
(UK only)
CEE 7/7*
220-230V Plug
IP-X0
*Other international plugs are available. Contact the manufacturer for more information.
**Domestic ground shipping information. Contact the manufacturer for export weight and dimensions.
EC1850BL
*Other international plugs are available, contact the manufacturer for more information.
WARNING
To prevent serious injury, death, or
property damage:
All electrical connections must be
made by a qualified and trained service
technician in accordance with applicable
electrical codes.
This appliance must be adequately
grounded in accordance with local
electrical codes or, in the absence of local
codes, with the current edition of the
National Electrical Code ANSI/NFPA No.
70. In Canada, all electrical connections
are to be made in accordance with CSA
C22.1, Canadian Electrical Code Part 1 or
local codes.
CE-approved appliances include an
equipotential-bonding terminal marked
with the symbol shown on the left .
Provisions for earthing are to be made in
accordance with IEC:2010 60335-1 section
27 or local codes.