15
MN-39362 •
Rev 0 • 04/18
• Titan Series
EC1850BL
Combination Blanket/Fluid Warmer
Operation Instructions
1 . Connect the warmer to an appropriate hospital grade
receptacle — specifi ed on the electrical information
page .
2 . Locate the circuit breaker on the back
of the warmer .
3 . Push the circuit breaker to the ON (I) position .
4 . Press the ON/OFF button
6
. The previous mode
of operation indicator
3
or
9
illuminates and the
temperature set-point is displayed
2
.
5 . Press the irrigation button
4
or the injection button
8
to select the desired mode of operation .
6 . Press the up arrow button
5
or down arrow button
7
to adjust the temperature set-point
2
.
NOTE:
Cool the chamber prior to changing from a higher
temperature setting to a lower temperature setting
to prevent an unwanted alarm
.
6 . Make sure the fl uid chamber insert is installed .
7 . Load the fl uid chamber with appropriate fl uid
containers .
8 . Make sure the warmer door is securely closed aft er
initial loading and following each fl uid removal .
Operating the Fluid Chamber
CAUTION
This appliance is not approved for
warming blood or blood products.
WARNING
Verify the fluid temperature prior to using the fluid.
Refer to fluid manufacturer’s labeling for recommended
warming procedures.
Injection fluid manufacturer suggests not to warm
injection fluids above 40°C (104°F). If fluids are warmed
above the suggested temperature, the fluids should
be discarded.
CAUTION
If the fluid warmer controller has failed, or if error
messages are displayed, it is recommended that the fluid
inventory be discarded. Refer to the troubleshooting
guide for error descriptions and action required.
1
7
8
9
10
11
2
4
3
5 6
NOTE:
Fluid chamber capacity is
30 1-liter bottles or
24-1-liter bags.