INVUE GV148 LIQUID CHEMICAL CONCENTRATION MONITOR
9
User Guide | Entegris, Inc.
2. For improved bubble rejection, mounting should
be in this orientation:
3. For maximum bubble rejection, mounting should
be in this orientation:
NOTE: This orientation is not recommended for
sediment carrying media.
Step 3: Connect Fluid Lines
1. Attach tubing using standard installation procedure
for Flaretek, PrimeLock, and Super Type 300 Pillar
connections.
NOTE: For detailed Flaretek and PrimeLock tube fitting
assembly instructions, visit
. For
detailed Super 300 Type Pillar assembly instructions,
contact Nippon Pillar Packaging Company, Ltd.
ELECTRICAL CONNECTIONS
Connect Wiring
NOTE: The wiring table below references the 19-pin PVDF
male Turck connector on the device (see image below). To
connect to the device, multiple cable options are available
with specific wiring diagrams. Please contact Entegris for
details.
PIN
FUNCTION
A
24 VDC power input
B
Power supply ground
C
TX-A
D
RX-A
E
Comms ground
F
TX-B
G
RX-B
H
Digital IO power 12-24 VDC
I
Digital output 0, system alarm
K
Digital output 1, future use
L
Digital input 0, RI re-zero
M
Digital input 1, Selection between
primary and secondary temperature
compensation coefficients
N
Digital IO ground
O
Analog output 1, concentration
P
Analog ground
R
Analog output 2, fluid temperature
S
Analog ground
T
Analog output 3, refractive index
U
Analog ground