INVUE GV148 LIQUID CHEMICAL CONCENTRATION MONITOR
User Guide | Entegris, Inc.
c. Enter the “Input: RI Value to Rezero”.
(In the case of DI water, enter 1.332987.)
d. Confirm monitor status is Ready before
pressing “Start”.
– The test will run for approximately 40 seconds
and is designed to calculate an offset between
the desired RI and actual measured RI.
e. Upon completion, the test will either show Pass
or Fail. The test field will fill in with statistics.