Check the protective conductor
4.7
Proceed as follows to check the protective conductor con-
nection between the generator and the consumer.
The following requirements must be met:
generator has been started (see 4.4)
device is connected (see 4.6)
device switched off
Fig. 4-4: Check the protective conductor
1.
Insert test cable into socket
(Fig. 4-4-(2))
.
2.
Hold a test tip an a metallic, blank location on the device.
The test lamp
(Fig. 4-4-(1))
on the multifunction display
shows the result:
Test lamp
Significance
lights up
green
protective conductor is OK
stays off
Protective conductor defective / not present
Table 4.1: Protective conductor test lamp
The protective conductor / potential equalization for this
device has been checked.
Requirements
Check the protective
conductor
Proceed as follows to
check the protective
conductor between the
consumer and the
generator: