Wiring
Stamoclean CAT430
16
Hauser
4
Wiring
4.1
Electrical connection
#
Warning!
• The electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable before beginning the connection work.
a0007284
Fig. 13:
Connection box: left cover, right inside the box
1
Transformer 24 V / 2A
2
Terminal block
3
4-pole connection
4
Fuse 1.6 A 5 x 20 mm (0.20 x 0.79 inch)
1.
First of all, remove the covers on the cable ducts (
13) and screw the connection box on.
2.
Disconnect the 4-pole connection between cover and terminal PCB (pos. 3) and secure the
cover.
3.
Guide the power cable through the Pg cable gland into the pump box. Pull the cable through
the right-hand cable duct into the connection box.
4.
Shorten the cable accordingly and connect the wires to terminals 15, 20 and 24.
5.
Pull both wires for the supply lead of the 24 V - hose heater through the left-hand cable duct
to the connection box and clamp the wires to terminals 1 and 6.
6.
Pull the supply lead of the orange thermal element on the electric band heater in the left-hand
cable duct up to the connection box.
7.
Shorten the cable accordingly and connect the wires to terminals 8, 14 and 23.
1
2
4
3
Power
Netz
Control
Steuerung
Evacuate
Entlüften
23
24
7
8
9
13 14 15
3
4
21 22
19
20
25 26
16 17 18
10 11 12
1
2
5
6