Installation
Stamoclean CAT430
10
Hauser
!
Note!
You will need a hammer drill with a 12 mm (0.47 inch) drill bit.
1.
Drill four holes with a diameter of 12 mm (0.47 inch) into the side of the wall facing the basin.
The distance from the upper end of the wall should be approx. 10 cm (3.94 inch).
2.
Screw the holders in.
Make sure, the basin holder is at least 60 cm (23.6 inch) below the water level (resp. the open
channel holder at least 35 cm (13.8 inch)).
3.3.2
Mounting the retaining slide and filtrate pipe
Perform the following steps outside the basin resp. the open channel:
1.
Feed the filtrate hose (
6, pos. 3) of the filtrate line (pos. 1) through the T-piece at the
PVC retaining slide (pos. 5).
2.
Screw the filtrate line in tight with the PVC hose gland of the T-piece.
3.
Fasten the stainless steel chain at the retaining slide by directing it through the heaxagon nut
(pos. 8).
4.
Remove the PVC screw caps (pos. 4) and place them so that they can be reached.
a0007156
Fig. 5:
Detail: wall mounting of the filter holder
(accessories)
a0007150
Fig. 6:
Filter element with retaining slide
1
2
3
4
5
6
7
8
Filtrate line
T-piece
Filtrate hose
PVC thread cap
Retaining slide
Filter element
Mounting guide pipe
Hexagon nut
150 (5.91)
180 (7.09)
Ø12
(0.47
)
mm (inch)