background image

Smartec S CLD134

Commissioning

Hauser

11

C124

Calculated cell
constant is displayed

0.1 ... 

6.3

 ... 9.99 cm

-1

6.300 1/cm
Cellconst

The calculated cell constant is displayed 
and entered in A5.

C125

Calibration status is 
displayed

o.k.
E xxx

o.k.
Status

If the calibration status is not o.k., the 
second display line shows an explanation 
of the error.

C126

Store calibration 
results

yes

no
new

yes
Store

If C125 = E xxx, then only no or 

new

.

If new, return to C.
If yes/no, return to “Measurement”.

C1(3)

Calibration with 
sensor adaptation for 
inductive sensors

Airs=airset (1)
Cellc= cell constant (2)

InstF=installation 
factor(3)

InstF
Calibrat

Sensor calibration with compensation of 
wall influence.
The distance from the sensor to the pipe 
wall and the pipe material (conductive or 
nonconductive) influence the measured 
value. The installation factor compensates 
this influence.

The sensor is installed in the process.

C131

Enter process 
temperature

25 °C

-35.0 ... 250.0 °C

25.0 °C
MTC temp.

Only exists if B1 = fixed.

C132

Enter 

α

 value of 

calibration solution

2.10 %K

0.00 ... 20.00 %K

2.10 %K
alpha value

This value is specified in the Technical 
Information of all E+H calibration 
solutions. You can also use the printed-on 
table to calculate the value.
Set 

α

 to 0 for calibration with 

uncompensated values.

C133

Enter correct 
conductivity value of 
calibration solution

current measured value

0.0 ... 9999 mS/cm

10.30 mS/cm
Real val.

Determine the correct conductivity value 
by a reference measurement.

C134

Calculated 
installation factor is 
displayed

1

0.10 ... 5.00

1
InstFact

The distance of the sensor to the pipe wall 
and the pipe material (conductive or 
nonconductive) influence the measured 
value. The installation factor compensates 
this influence.

C135

Calibration status is 
displayed

o.k.
E xxx

o.k.
Status

If the calibration status is not o.k., the 
second display line shows an explanation 
of the error.

C136

Store calibration 
results?

yes

no
new

yes
Store

If C135 = E xxx, then only no or 

new

.

If new, return to C.
If yes/no, return to “Measurement”.

Code

Field

Selection or range

Display

Info

Содержание Smartec S CLD134

Страница 1: ...structions Smartec S CLD134 Conductivity Measuring System These instructions are Brief Operating Instructions For detailed information please read the Operating Instructions The complete device docume...

Страница 2: ...d media It is particularly suitable for use in the foodstuffs industry Any other use than the one described here compromises the safety of persons and the entire measuring system and is therefore not...

Страница 3: ...te version Wall mounting For wall mounting attach the mounting plate to the wall by drilling holes as required Anchors and screws are to be provided by the operator Note Wall mounting is not recommend...

Страница 4: ...ng Warning The electrical connection must only be carried out by a certified electrician Technical personnel must have read and understood the instructions in this manual and must adhere to them Ensur...

Страница 5: ...tput 2 temperature Auxiliary power output Binary input 2 MRS 1 2 Binary input 1 hold MRS 3 4 F G H I MRS Conductivity sensor Temperature sensor Alarm contact position no current Power supply Remote pa...

Страница 6: ...not energized S2 External contacts not energized 3 3 Structure and termination of measuring cable 85 S1 D1 15 V A B D2 86 81 82 93 94 S2 a0006037 en Fig 5 Structure of the measuting cable a0006036 Fi...

Страница 7: ...ve on within function groups and In setup mode the PLUS and MINUS keys have the following functions Selection of function groups Setting of parameters and numeric values In measuring mode repeatedly p...

Страница 8: ...simultaneously to quit the Service function group 7 Press the MINUS key several times until the Setup 1 function group is displayed 8 Press the ENTER key to edit Setup 1 SETUP 1 9 In A1 select the op...

Страница 9: ...eal temperature 35 0 250 0 C 0 0 C RealTemp 22 The difference between the measured and the entered temperatures is displayed Press the ENTER key The display returns to the initial display of the Setup...

Страница 10: ...not o k the second display line shows an explanation of the error C114 Store calibration results yes no new yes Store If C113 E xxx then only no or new If new return to C If yes no return to Measurem...

Страница 11: ...mperature 25 C 35 0 250 0 C 25 0 C MTC temp Only exists if B1 fixed C132 Enter value of calibration solution 2 10 K 0 00 20 00 K 2 10 K alpha value This value is specified in the Technical Information...

Страница 12: ...www endress com worldwide KA401C 07 en 07 07 Printed in Germany FM SGML 6 0 71059831...

Отзывы: