background image

Wiring

Smartec S CLD134

6

Hauser

Connection of binary inputs

a0005639

Fig. 4:  Connection of binary inputs when using external contacts

A

Auxiliary power output

B

Contact inputs D1 and D2

S1

External contacts, not energized

S2

External contacts, not energized

3.3

Structure and termination of measuring cable

85

S1

D1

15 V

A

B

D2

86

81

82

93

94

S2

a0006037-en

Fig. 5:  Structure of the measuting cable

a0006036

Fig. 6:  Electrical connection of the CLS54 sensor for the 

separate version

(84)

(83)

GN (11)

WH (12)

YE (13)

CLK5

RD

WH

Semiconductor

screen

Screen

(16)

Screen

(15)

BN

(unused)

J

83

11

S

12

16 15 84

CLK5

13

GN WH YE

RD

WH

BU

RD

RD

BU

Содержание Smartec S CLD134

Страница 1: ...structions Smartec S CLD134 Conductivity Measuring System These instructions are Brief Operating Instructions For detailed information please read the Operating Instructions The complete device docume...

Страница 2: ...d media It is particularly suitable for use in the foodstuffs industry Any other use than the one described here compromises the safety of persons and the entire measuring system and is therefore not...

Страница 3: ...te version Wall mounting For wall mounting attach the mounting plate to the wall by drilling holes as required Anchors and screws are to be provided by the operator Note Wall mounting is not recommend...

Страница 4: ...ng Warning The electrical connection must only be carried out by a certified electrician Technical personnel must have read and understood the instructions in this manual and must adhere to them Ensur...

Страница 5: ...tput 2 temperature Auxiliary power output Binary input 2 MRS 1 2 Binary input 1 hold MRS 3 4 F G H I MRS Conductivity sensor Temperature sensor Alarm contact position no current Power supply Remote pa...

Страница 6: ...not energized S2 External contacts not energized 3 3 Structure and termination of measuring cable 85 S1 D1 15 V A B D2 86 81 82 93 94 S2 a0006037 en Fig 5 Structure of the measuting cable a0006036 Fi...

Страница 7: ...ve on within function groups and In setup mode the PLUS and MINUS keys have the following functions Selection of function groups Setting of parameters and numeric values In measuring mode repeatedly p...

Страница 8: ...simultaneously to quit the Service function group 7 Press the MINUS key several times until the Setup 1 function group is displayed 8 Press the ENTER key to edit Setup 1 SETUP 1 9 In A1 select the op...

Страница 9: ...eal temperature 35 0 250 0 C 0 0 C RealTemp 22 The difference between the measured and the entered temperatures is displayed Press the ENTER key The display returns to the initial display of the Setup...

Страница 10: ...not o k the second display line shows an explanation of the error C114 Store calibration results yes no new yes Store If C113 E xxx then only no or new If new return to C If yes no return to Measurem...

Страница 11: ...mperature 25 C 35 0 250 0 C 25 0 C MTC temp Only exists if B1 fixed C132 Enter value of calibration solution 2 10 K 0 00 20 00 K 2 10 K alpha value This value is specified in the Technical Information...

Страница 12: ...www endress com worldwide KA401C 07 en 07 07 Printed in Germany FM SGML 6 0 71059831...

Отзывы: