background image

Commissioning

Smartec S CLD134

10

Hauser

!

Note! 
You must perform an air set before installing the sensor. To do so, refer to the chapter 
"Calibration".

4.4

Calibration

To access the "Calibration" function group, press the CAL key (code = 22).

Code

Field

Selection or range

Display

Info

C

Function group
CALIBRATION

CALIBRAT

Calibration settings

C1(1)

Compensation of 
residual coupling

Airs=Airset (1)

Cellc=cell constant (2)
InstF=Installation factor(3)

AirS
Calibrat

When commissioning inductive sensors, 
an airset is mandatory.
The calibration of the sensor is to be 
performed in air. The sensor must be dry.

Remove sensor from the medium and dry 

completely

.

C111

Residual coupling
start calibration
(airset)

current measured value

0.0 

μ

S/cm

AirSet

Start calibration with CAL.

C112

Residual coupling is 
displayed

-80.0 ... 80.0 

μ

S/cm

5.3 

μ

S/cm

AirSetVal

Residual coupling of measuring system 
(sensor and transmitter).

C113

Calibration status is 
displayed

o.k.
E xxx

o.k.
Status

If the calibration status is not o.k., the 
second display line shows an explanation 
of the error.

C114

Store calibration 
results

yes

no
new

yes
Store

If C113 = E xxx, then only no or 

new

.

If new, return to C.
If yes/no, return to “Measurement”.

C1(2)

Calibration of cell 
constant

Airs=Airset (1)

Cellc=cell constant (2)

InstF=installation factor(3)

Cellc
Calibrat

Immerse sensor in calibration solution.

!

Note! 

This section decribes the calibration for temperature compensated 
conductivity. For calibration with uncompensated conductivity, set the 
temperature coefficient 

α

 to 0.

The sensor should be immersed at a 
sufficient distance from the vessel wall 
(installation factor has no influence if 
a > 15 mm / 0.59").

C121

Enter process
temperature (MTC)

25 °C

-35.0 ... 250.0 °C

25.0 °C
ProcTemp.

Only exists if B1 = fixed

C122

Enter 

α

 value of 

calibration solution

2.10 %K

0.00 ... 20.00 %K

2.10 %K
alpha val

This value is specified in the Technical 
Information of all E+H calibration 
solutions. You can also use the printed-on 
table to calculate the value.
Set 

α

 to 0 for calibration with 

uncompensated values.

C123

Enter correct 
conductivity value of 
calibration solution

current measured value

0.0 ... 9999 mS/cm

10.30 mS/cm
Real. val

The display is always in mS/cm.

Содержание Smartec S CLD134

Страница 1: ...structions Smartec S CLD134 Conductivity Measuring System These instructions are Brief Operating Instructions For detailed information please read the Operating Instructions The complete device docume...

Страница 2: ...d media It is particularly suitable for use in the foodstuffs industry Any other use than the one described here compromises the safety of persons and the entire measuring system and is therefore not...

Страница 3: ...te version Wall mounting For wall mounting attach the mounting plate to the wall by drilling holes as required Anchors and screws are to be provided by the operator Note Wall mounting is not recommend...

Страница 4: ...ng Warning The electrical connection must only be carried out by a certified electrician Technical personnel must have read and understood the instructions in this manual and must adhere to them Ensur...

Страница 5: ...tput 2 temperature Auxiliary power output Binary input 2 MRS 1 2 Binary input 1 hold MRS 3 4 F G H I MRS Conductivity sensor Temperature sensor Alarm contact position no current Power supply Remote pa...

Страница 6: ...not energized S2 External contacts not energized 3 3 Structure and termination of measuring cable 85 S1 D1 15 V A B D2 86 81 82 93 94 S2 a0006037 en Fig 5 Structure of the measuting cable a0006036 Fi...

Страница 7: ...ve on within function groups and In setup mode the PLUS and MINUS keys have the following functions Selection of function groups Setting of parameters and numeric values In measuring mode repeatedly p...

Страница 8: ...simultaneously to quit the Service function group 7 Press the MINUS key several times until the Setup 1 function group is displayed 8 Press the ENTER key to edit Setup 1 SETUP 1 9 In A1 select the op...

Страница 9: ...eal temperature 35 0 250 0 C 0 0 C RealTemp 22 The difference between the measured and the entered temperatures is displayed Press the ENTER key The display returns to the initial display of the Setup...

Страница 10: ...not o k the second display line shows an explanation of the error C114 Store calibration results yes no new yes Store If C113 E xxx then only no or new If new return to C If yes no return to Measurem...

Страница 11: ...mperature 25 C 35 0 250 0 C 25 0 C MTC temp Only exists if B1 fixed C132 Enter value of calibration solution 2 10 K 0 00 20 00 K 2 10 K alpha value This value is specified in the Technical Information...

Страница 12: ...www endress com worldwide KA401C 07 en 07 07 Printed in Germany FM SGML 6 0 71059831...

Отзывы: