Smartec S CLD134

Commissioning

Hauser

9

15. If you are working in applications that fluctuate a great deal and 

you need to stabilize the display, enter the required damping 
factor in A7.
Confirm your entry by pressing ENTER.
The display returns to the inital display of "Setup 1".

1

1 ... 60

1
Damping

16. Press the MINUS key to go to the "Setup 2" function group.

17. Press the ENTER key to edit "Setup 2".

SETUP 2

18. In B1, select the temperature sensor of your conductivity sensor. 

By default, your measuring system is supplied with the CLS54 
sensor with Pt 1000 temperature sensor.
Confirm your entry by pressing ENTER.

Pt100

Pt1k = Pt 1000

NTC30
fixed

Pt1k
ProcTemp.

19. In B2, select the appropriate temperature compensation for your 

process, e.g. "lin" = linear.
Confirm your selection by pressing ENTER.

none

lin = linear

NaCl = common salt 
(IEC 60746)
Tab 1 ... 4

lin
TempComp.

20. In B3, enter the temperature coefficient 

α

.

Confirm your entry by pressing ENTER.

2.1 %/K

0.0 ... 20.0 %/K

2.10 %/K
Alpha val

21. The real temperature is displayed in B5. If necessary, calibrate the 

temperature sensor to an external measurement.
Confirm your entry by pressing ENTER.

Display and entry of real 
temperature
-35.0 ... 250.0 °C

0.0 °C
RealTemp.

22. The difference between the measured and the entered 

temperatures is displayed.
Press the ENTER key. 
The display returns to the initial display of the "Setup 2" function 
group.

0.0 °C

-5.0 ... 5.0 °C

0.0 °C
TempOffs.

23. Press the MINUS key to go to the "Output" function group.

24. Press the ENTER key to edit the output settings.

OUTPUT

25. In O1, select your output, e.g. "out1" = output 1.

Confirm your selection by pressing ENTER.

out 1

out 2

out1
Sel. Out

26. In O2, select the linear characteristic.

Confirm your selection by pressing ENTER.

lin = linear (1)

sim = simulation (2)

lin
Sel. Type

27. In O211, select the current range for your output, 

e.g. 4 ... 20 mA.
Confirm your selection by pressing ENTER.

4 ... 20 mA

0 ... 20 mA

4-20
Sel.Range

28. In O212, enter the conductivity corresponding to the minium 

current value at the transmitter output, e.g. 0

μ

S/cm.

Confirm your entry by pressing ENTER.

0.00 

μ

S/cm

0.00 

μ

S/cm ... 

2000 mS/cm

μ

S/cm

0/4 mA

29. In O213, enter the conductivity corresponding to the maximum 

current value at the transmitter output, e.g. 930 mS/cm.
Confirm your entry by pressing ENTER.
The display returns to the initial display of the "Output" function 
group.

2000 mS/cm

0.0 

μ

S/cm ... 2000 mS/cm

930 mS/cm
20 mA

30. Press the PLUS and MINUS keys simultaneously to return to 

measuring mode.

Input

Selection or range
(factory setting bold)

Display

Содержание Smartec S CLD134

Страница 1: ...structions Smartec S CLD134 Conductivity Measuring System These instructions are Brief Operating Instructions For detailed information please read the Operating Instructions The complete device docume...

Страница 2: ...d media It is particularly suitable for use in the foodstuffs industry Any other use than the one described here compromises the safety of persons and the entire measuring system and is therefore not...

Страница 3: ...te version Wall mounting For wall mounting attach the mounting plate to the wall by drilling holes as required Anchors and screws are to be provided by the operator Note Wall mounting is not recommend...

Страница 4: ...ng Warning The electrical connection must only be carried out by a certified electrician Technical personnel must have read and understood the instructions in this manual and must adhere to them Ensur...

Страница 5: ...tput 2 temperature Auxiliary power output Binary input 2 MRS 1 2 Binary input 1 hold MRS 3 4 F G H I MRS Conductivity sensor Temperature sensor Alarm contact position no current Power supply Remote pa...

Страница 6: ...not energized S2 External contacts not energized 3 3 Structure and termination of measuring cable 85 S1 D1 15 V A B D2 86 81 82 93 94 S2 a0006037 en Fig 5 Structure of the measuting cable a0006036 Fi...

Страница 7: ...ve on within function groups and In setup mode the PLUS and MINUS keys have the following functions Selection of function groups Setting of parameters and numeric values In measuring mode repeatedly p...

Страница 8: ...simultaneously to quit the Service function group 7 Press the MINUS key several times until the Setup 1 function group is displayed 8 Press the ENTER key to edit Setup 1 SETUP 1 9 In A1 select the op...

Страница 9: ...eal temperature 35 0 250 0 C 0 0 C RealTemp 22 The difference between the measured and the entered temperatures is displayed Press the ENTER key The display returns to the initial display of the Setup...

Страница 10: ...not o k the second display line shows an explanation of the error C114 Store calibration results yes no new yes Store If C113 E xxx then only no or new If new return to C If yes no return to Measurem...

Страница 11: ...mperature 25 C 35 0 250 0 C 25 0 C MTC temp Only exists if B1 fixed C132 Enter value of calibration solution 2 10 K 0 00 20 00 K 2 10 K alpha value This value is specified in the Technical Information...

Страница 12: ...www endress com worldwide KA401C 07 en 07 07 Printed in Germany FM SGML 6 0 71059831...

Отзывы: