Commissioning
Smartec S CLD134
8
Hauser
4.3
Quick setup
After switching the transmitter on, configure the major functions required for accurate
measurement. The following section gives you an example for a basic configuration.
Input
Selection or range
(factory setting bold)
Display
1.
Press the ENTER key.
2.
Enter the code 22 to be able to edit the setup. Press the ENTER
key.
3.
Press the MINUS key several times until the "Service" function
group is displayed.
4.
Press the ENTER key to edit this function group.
SERVICE
5.
Select your language, e.g. "ENG" for English.
Confirm your entry by pressing the ENTER key.
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
ENG
Language
6.
Press the PLUS and MINUS keys simultaneously to quit the
"Service" function group.
7.
Press the MINUS key several times until the "Setup 1" function
group is displayed.
8.
Press the ENTER key to edit "Setup 1".
SETUP 1
9.
In A1, select the operating mode, e.g. "cond" = conductivity.
Confirm your selection by pressing the ENTER key.
cond = conductivity
conc = concentration
cond
Oper.Mode
10. In A2, press the ENTER key to confirm the factory setting.
%
ppm
mg/l
TDS = Total Dissolved
Solids
none
ppm
Conc.Unit
11. In A3, press the ENTER key to confirm the factory setting.
XX.xx
X.xxx
XXX.x
XXXX
XX.xx
Format
12. In A4, press the ENTER key to confirm the factory settting.
auto
,
μ
S/cm, mS/cm,
S/cm,
μ
S/m, mS/m, S/m
auto
Unit
13. In A5, enter the cell constant for the connected sensor. Refer to
the sensor’s or the compact version’s quality certificate for the
exact value.
0.10 ...
6.3
... 9.99
6.300 l/cm
Cellconst
14. In A6, press the ENTER key to confirm the factory setting.
0.10 ...
1
... 5.00
1.000
InstFac