Commissioning

Smartec S CLD134

8

Hauser

4.3

Quick setup

After switching the transmitter on, configure the major functions required for accurate 
measurement. The following section gives you an example for a basic configuration.

Input

Selection or range
(factory setting bold)

Display

1.

Press the ENTER key.

2.

Enter the code 22 to be able to edit the setup. Press the ENTER 
key.

3.

Press the MINUS key several times until the "Service" function 
group is displayed.

4.

Press the ENTER key to edit this function group.

SERVICE

5.

Select your language, e.g. "ENG" for English.
Confirm your entry by pressing the ENTER key.

ENG = English

GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish

ENG
Language

6.

Press the PLUS and MINUS keys simultaneously to quit the 
"Service" function group.

7.

Press the MINUS key several times until the "Setup 1" function 
group is displayed.

8.

Press the ENTER key to edit "Setup 1".

SETUP 1

9.

In A1, select the operating mode, e.g. "cond" = conductivity.
Confirm your selection by pressing the ENTER key.

cond = conductivity

conc = concentration

cond
Oper.Mode

10. In A2, press the ENTER key to confirm the factory setting.

%

ppm

mg/l
TDS = Total Dissolved 
Solids
none

ppm
Conc.Unit

11. In A3, press the ENTER key to confirm the factory setting.

XX.xx

X.xxx
XXX.x
XXXX

XX.xx
Format

12. In A4, press the ENTER key to confirm the factory settting.

auto

μ

S/cm, mS/cm, 

S/cm, 

μ

S/m, mS/m, S/m

auto
Unit

13. In A5, enter the cell constant for the connected sensor. Refer to 

the sensor’s or the compact version’s quality certificate for the 
exact value.

0.10 ... 

6.3

 ... 9.99

6.300 l/cm
Cellconst

14. In A6, press the ENTER key to confirm the factory setting.

0.10 ... 

1

 ... 5.00

1.000
InstFac

Содержание Smartec S CLD134

Страница 1: ...structions Smartec S CLD134 Conductivity Measuring System These instructions are Brief Operating Instructions For detailed information please read the Operating Instructions The complete device docume...

Страница 2: ...d media It is particularly suitable for use in the foodstuffs industry Any other use than the one described here compromises the safety of persons and the entire measuring system and is therefore not...

Страница 3: ...te version Wall mounting For wall mounting attach the mounting plate to the wall by drilling holes as required Anchors and screws are to be provided by the operator Note Wall mounting is not recommend...

Страница 4: ...ng Warning The electrical connection must only be carried out by a certified electrician Technical personnel must have read and understood the instructions in this manual and must adhere to them Ensur...

Страница 5: ...tput 2 temperature Auxiliary power output Binary input 2 MRS 1 2 Binary input 1 hold MRS 3 4 F G H I MRS Conductivity sensor Temperature sensor Alarm contact position no current Power supply Remote pa...

Страница 6: ...not energized S2 External contacts not energized 3 3 Structure and termination of measuring cable 85 S1 D1 15 V A B D2 86 81 82 93 94 S2 a0006037 en Fig 5 Structure of the measuting cable a0006036 Fi...

Страница 7: ...ve on within function groups and In setup mode the PLUS and MINUS keys have the following functions Selection of function groups Setting of parameters and numeric values In measuring mode repeatedly p...

Страница 8: ...simultaneously to quit the Service function group 7 Press the MINUS key several times until the Setup 1 function group is displayed 8 Press the ENTER key to edit Setup 1 SETUP 1 9 In A1 select the op...

Страница 9: ...eal temperature 35 0 250 0 C 0 0 C RealTemp 22 The difference between the measured and the entered temperatures is displayed Press the ENTER key The display returns to the initial display of the Setup...

Страница 10: ...not o k the second display line shows an explanation of the error C114 Store calibration results yes no new yes Store If C113 E xxx then only no or new If new return to C If yes no return to Measurem...

Страница 11: ...mperature 25 C 35 0 250 0 C 25 0 C MTC temp Only exists if B1 fixed C132 Enter value of calibration solution 2 10 K 0 00 20 00 K 2 10 K alpha value This value is specified in the Technical Information...

Страница 12: ...www endress com worldwide KA401C 07 en 07 07 Printed in Germany FM SGML 6 0 71059831...

Отзывы: