Appendix
RID14
44
Hauser
Navigation
Operation → "Display value X" submenu → Description
Description
Possibility to enter a user-defined text with a maximum of 16 characters.
This text is shown below the value on the display.
If the text is longer than 5 characters, it is displayed as marquee text.
User entry
Free text
Factory setting
Additional information
• Visible offline: Yes
• Writable offline: Yes
Enable % symbol
Navigation
Operation → "Display value X" submenu → Enable % symbol
Description
Switches on the %-symbol of the display.
Options
• On
• Off
Factory setting
Off
Additional information
• Visible offline: Yes
• Writable offline: Yes
Bargraph 0%
Navigation
Operation → "Display value X" submenu → Bargraph 0%
Description
Enter the minimum value (0%) for the bargraph.
Only visible if data type = analog and bar graph on/off = on (Expert menu).
User entry
Numerical value
Factory setting
0
Additional information
• Visible offline: Yes
• Writable offline: Yes
Bargraph 100%
Navigation
Operation → "Display value X" submenu → Bargraph 100%
Description
Enter the maximum value (100%) for the bargraph.
Only visible if data type = analog and bar graph on/off = on (Expert menu).
User entry
Numerical value
Factory setting
100
Additional information
• Visible offline: Yes
• Writable offline: Yes
Decimal places
Navigation
Operation → "Display value X" submenu → Decimal places
Содержание RID14