Operating the field indicator
RID14
26
Hauser
To set the DIP switches, proceed as follows:
1. Remove the housing cover and remove the display → 5, 13
2. Configure the DIP switches. Switch to ON = function enabled, switch to OFF =
function disabled.
3. Fit the display onto the electronics.
4. Close the housing cover and secure it.
1
ON
OFF
A0021499
14
Hardware configuration of the field indicator
ON
OFF
1
2
4
8
16 32 64
Data
Address
Set
AI1
AI2
DIP
Write
protect
Remote
Offset
A0024593-EN
15
Assignment of the DIP switches
6.5.1
Switching write protection on/off
Write protection is switched on and off via the "WRITE LOCK" DIP switch. When write
protection is active ("WRITE LOCK" is "ON"), parameters cannot be modified. The current
write protection status is displayed in the "Locking status" parameter. When write
protection is active ("WRITE LOCK" is "ON"), a padlock symbol is lit on the display.
6.5.2
Choosing between operation via DIP switches and remote
operation
When the device is operated via the DIP switches only 2 values are displayed even if
more display values were configured previously in the configuration software.
The bargraph is not displayed when the device is operated via the DIP switches.
Via the "Remote/DIP" DIP switch users can specify whether configuration is possible onsite
via the DIP switches or remotely via the DTM and PC configuration software. If the switch
is set to "OFF" (remote), all the switches apart from "WRITE LOCK" are disabled. If the
switch is set to "ON", all the DIP switches work and operation via the DTM is not possible.
6.5.3
Setting the bus address
The DIP switches can be used to set the address of the Profibus PA measuring device
whose values are to be displayed on the field indicator.
The bus address is set as follows:
1. Use DIP switch "AI1/AI2" to select whether the configured address refers to analog
input 1 (switch set to "ON") or analog input 2 (switch set to "OFF").
Содержание RID14