Proline Promass E 300 HART
Diagnostics and troubleshooting
Hauser
155
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
372
Sensor electronic (ISEM)
faulty
1. Restart device
2. Check if failure recurs
3. Replace sensor electronic module
(ISEM)
F
Alarm
373
Sensor electronic (ISEM)
faulty
1. Transfer data or reset device
2. Contact service
F
Alarm
374
Sensor electronic (ISEM)
faulty
1. Restart device
2. Check if failure recurs
3. Replace sensor electronic module
(ISEM)
S
Warning
375
I/O- 1 to n
communication failed
1. Restart device
2. Check if failure recurs
3. Replace module rack inclusive
electronic modules
F
Alarm
378
Supply voltage ISEM
faulty
Check supply voltage to the ISEM
F
Alarm
382
Data storage
1. Insert T-DAT
2. Replace T-DAT
F
Alarm
383
Memory content
1. Restart device
2. Delete T-DAT via 'Reset device'
parameter
3. Replace T-DAT
F
Alarm
387
HistoROM data faulty
Contact service organization
F
Alarm
Diagnostic of configuration
330
Flash file invalid
1. Update firmware of device
2. Restart device
M
Warning
331
Firmware update failed
1. Update firmware of device
2. Restart device
F
Warning
410
Data transfer
1. Check connection
2. Retry data transfer
F
Alarm
412
Processing download
Download active, please wait
C
Warning
431
Trim 1 to n
Carry out trim
C
Warning
437
Configuration
incompatible
1. Restart device
2. Contact service
F
Alarm
438
Dataset
1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
M
Warning
441
Current output 1 to n
1. Check process
2. Check current output settings
S
Warning
442
Frequency output 1 to n
1. Check process
2. Check frequency output settings
S
Warning
443
Pulse output 1 to n
1. Check process
2. Check pulse output settings
S
Warning
444
Current input 1 to n
1. Check process
2. Check current input settings
S
Warning
453
Flow override
Deactivate flow override
C
Warning
484
Failure mode simulation Deactivate simulation
C
Alarm
485
Measured variable
simulation
Deactivate simulation
C
Warning