Technical data
Proline Promass E 300 HART
180
Hauser
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronics temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
Switch output
Maximum input values
DC 30 V, 250 mA (passive)
Open-circuit voltage
DC 28.8 V (active)
Switching behavior
Binary, conductive or non-conductive
Switching delay
Configurable: 0 to 100 s
Number of switching
cycles
Unlimited
Assignable functions
• Off
• On
• Diagnostic behavior
• Limit value
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Totalizer 1-3
• Flow direction monitoring
• Status
• Partially filled pipe detection
• Low flow cut off
The range of options increases if the measuring device has one or more
application packages.
Double pulse output
Function
Double pulse
Version
Open collector
Can be set to:
• Active
• Passive
• Passive NAMUR
Maximum input values
DC 30 V, 250 mA (passive)
Open-circuit voltage
DC 28.8 V (active)
Voltage drop
For 22.5 mA: ≤ DC 2 V
Output frequency
Configurable: 0 to 1
000 Hz
Damping
Configurable: 0 to 999 s