Proline Promass 80H, 83H
Hauser
13
Zero point stability:
Sample calculation
a0004611
Max. measured error in % of measured value (example: Promass 80H, 83H/ DN 25)
Calculation example (mass flow, liquid):
Given: Promass 83 H / DN 25, flow measured value = 8000 kg/h
Max. measured error: ±0.125% ± [(zero point stability / measured value)
⋅
x 100]% o.r.
Max. measured error: ±0.125% ±1.8 kg/h : 8000 kg/h
⋅
100% = ±0.15%
Density (liquid)
1 g/cc = 1 kg/l
Standard calibration:
±0.02 g/cc
Special density calibration (optional), calibration range: 0.8 to 1.8 g/cc, 5 to 80 °C:
±0.002 g/cc
After field density calibration or under reference conditions:
±0.0010 g/cc
Temperature
±0.5 °C ±0.005
⋅
T (T = fluid temperature in °C)
DN
Maximum full scale value
Zero point stability
[kg/h]
[kg/h]
8
2000
0.20
15
6500
0.65
25
18000
1.8
40
45000
4.5
50
70000
7.0
0
±0.5
±0.2
±1.0
[%]
0
1
2
4
6
8
10
12
14
16
18
t/h
Promass 83
Promass 80