Proline Promass 80H, 83H
12
Hauser
Connecting cable for remote version:
• Cable entry M20 x 1.5 (8 to 12 mm)
• Thread for cable entries, 1/2" NPT, G 1/2"
Cable specification
Remote version
• 6 x 0.38 mm
2
(PVC cable with common shield and individually shielded cores
• Conductor resistance:
≤
50
Ω
/km
• Capacitance: core/shield:
≤
420 pF/m
• Cable length: max. 20 m
• Permanent operating temperature: max. +105 °C
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC
requirements of EN 61326/A1, and NAMUR recommendation NE 21/43.
Power consumption
AC: <15 VA (including sensor)
DC: <15 W (including sensor)
Switch-on current
:
• Max. 13.5 A (< 50 ms) at 24 V DC
• Max. 3 A (< 5 ms) at 260 V AC
Power supply failure
Promass 80
Lasting min. 1 power cycle:
• EEPROM saves measuring system data if the power supply fails
• HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial
number, calibration factor, zero point, etc.)
Promass 83
Lasting min. 1 power cycle:
• EEPROM and T-DAT save the measuring system data if the power supply fails.
• HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial
number, calibration factor, zero point, etc.)
Potential equalization
No measures necessary.
Performance characteristics
Reference operating
conditions
Error limits following ISO/DIS 11631:
• 20 to 30 °C
• 2 to 4 bar
• Calibration systems as per national norms
• Zero point calibrated under operating conditions
• Field density calibrated (or special density calibration)
Maximum measured error
The following values refer to the pulse/frequency output. Measured error at the current output is typically
±5 A.
o.r. = of reading
Mass flow (liquid):
Promass 80:
±0.175% ± [(zero point stability / measured value)
⋅
100]% o.r.
Promass 83:
±0.125% ± [(zero point stability / measured value)
⋅
100]% o.r.
Volume flow (liquid)
±0.50% ± [(zero point stability / measured value)
⋅
100]% o.r.