Proline Promass 80H, 83H
Hauser
19
Operating conditions: Environment
Ambient temperature range
Standard: –20 to +60 °C (sensor, transmitter)
Optional: –40 to +60 °C (sensor, transmitter)
!
Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C the readability of the display may be impaired.
Storage temperature
–40 to +80 °C (pref20 °C)
Degree of protection
Standard: IP 67 (NEMA 4X) for transmitter and sensor
Shock resistance
According to IEC 68-2-31
Vibration resistance
Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6
Electromagnetic compatibility
(EMC)
As per EN 61326/A1 (IEC 1326) and NAMUR recommendation NE 21
Operating conditions: Process
Fluid temperature range
Sensor
–50 to +200 °C
Fluid pressure range
(nominal pressure)
Flanges:
DIN PN 40 / ASME B16.5 Cl 150, Cl 300 / JIS 10K, 20K
Pressure ranges of secondary containment:
DN 8 to 15: 25 bar
DN 25 to 50: 16 bar
#
Warning!
In case a danger of measuring tube failure exists due to process characteristics, e.g. with corrosive process
fluids, we recommend the use of sensors whose secondary containment is equipped with special pressure
monitoring connections (ordering option). With the help of these connections, fluid collected in the secondary
containment in the event of tube failure can be bled off. Dimensions
→
Page 25 ff.
Limiting flow
See information in the "Measuring range" section
→
Page 5
Select nominal diameter by optimizing between required flow range and permissible pressure loss. See the
"Measuring range" section for a list of maximum possible full scale values.
• The minimum recommended full scale value is approx. 1/20 of the max. full scale value.
• In most applications, 20 to 50% of the maximum full scale value can be considered ideal
• Select a lower full scale value for abrasive substances such as fluids with entrained solids (flow velocity
<1 m/s).