Proline Promass 80H, 83H
Hauser
5
Input
Measured variable
• Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
• Fluid density (proportional to resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)
Measuring range
Measuring ranges for liquids
F
a0003673
• Universal sensor for fluid temperatures up to 200 °C.
• Nominal diameters DN 8 to 250
• Tube material: stainless steel or Alloy C-22
Documentation No.
TI 053D/06/en
F (High-temperature)
a0003675
• Universal high-temperature sensor for fluid temperatures
up to 350 °C.
• Nominal diameters DN 25, 50, 80
• Tube material: Alloy C-22
M
a0003676
• Robust sensor for extreme process pressures, high
requirements for the secondary containment and fluid
temperatures up to 150 °C
• Nominal diameters DN 8 to 80
• Tube material: titanium
I
a0003678
• Straight single-tube instrument. Minimal shear stress on
fluid, hygienic design, low pressure loss.
• Nominal diameters DN 8 to 80
• Tube material: titanium
Documentation No.
TI 075D/06/en
S
a0006828
• Single bent tube.
Hygienic design, low pressure loss,
for fluid temperatures up to 150 °C
• Nominal diameters DN 8 to 50
• Tube material: stainless steel
Documentation No.
TI 076D/06/en
DN
Range for full scale values (liquids)
g
min(F)
to
g
max(F)
8
0 to 2000 kg/h
15
0 to 6500 kg/h
25
0 to 18000 kg/h
40
0 to 45000 kg/h
50
0 to 70000 kg/h